MULTIDETECTION IN ULTRAVIOLET ANGLE-RESOLVED PHOTOEMISSION

被引:3
作者
LEVEQUE, G
VILLACHONRENARD, Y
BONNET, J
机构
[1] UNIV MONTPELLIER 2,LEOCM,F-34095 MONTPELLIER,FRANCE
[2] UNIV PARIS 11,LURE,F-91405 ORSAY,FRANCE
关键词
D O I
10.1063/1.1143012
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have tested at Super ACO (LURE) a new angular resolved photoemission device for the 12-30 eV energy range, using multidetection of photoelectrons: all all electrons emitted in a plane are collected and focused onto a fluorescent screen in a coordinate grid directly related to (E(i), k parallel-to). This multidetection provides a substantial diminution of the acquisition time. Ten seconds are needed to obtain a complete off-normal set of energy distribution curves (EDC). Real time visualization of the valence-band structure for one azimuthal plane is even possible. Digital image treatment is performed after the experiments, permitting an easier extraction of the characteristics of the bands and the emission structures.
引用
收藏
页码:1525 / 1527
页数:3
相关论文
共 4 条
[1]   REALIZATION OF AN ELECTRON SPECTROMETER USING A TOROIDAL MAGNETIC DEFLECTOR [J].
LEVEQUE, G ;
ROBIN, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (08) :1456-1459
[3]   VALENCE BAND-STRUCTURE OF CD0.7HG0.3TE DETERMINED BY ANGULAR RESOLVED PHOTOEMISSION [J].
LEVEQUE, G ;
ORSAL, B ;
ALABEDRA, R ;
FLACHET, JC .
JOURNAL OF CRYSTAL GROWTH, 1990, 101 (1-4) :364-367
[4]   VALENCE-BAND AND SURFACE ELECTRONIC-STRUCTURE OF CDTE [J].
MAGNUSSON, KO ;
FLODSTROM, SA .
PHYSICAL REVIEW B, 1988, 38 (08) :5384-5391