THE DOSE DISTRIBUTION PRODUCED BY A P-32 COATED STENT

被引:44
作者
PRESTWICH, WV
KENNETT, TJ
KUS, FW
机构
[1] MCMASTER UNIV,COMP SERV,HAMILTON,ON L8S 4M1,CANADA
[2] MCMASTER UNIV,INFORMAT SERV,HAMILTON,ON L8S 4M1,CANADA
关键词
STENT; P-32; DOSE; RESTENOSIS;
D O I
10.1118/1.597455
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dose distribution surrounding uniform cylinders of 32P has been calculated in order to obtain information relevant to the utilization of radioactive stents. The radiological impact on two major components of vascular tissue, smooth muscle, and endothelial cells is also considered. © 1995, American Association of Physicists in Medicine. All rights reserved.