REFLECTANCE PULSE OXIMETRY MEASUREMENTS FROM THE RETINAL FUNDUS

被引:22
作者
DEKOCK, JP
TARASSENKO, L
GLYNN, CJ
HILL, AR
机构
[1] UNIV OXFORD,DEPT ENGN SCI,MED ENGN UNIT,OXFORD OX1 3PJ,OXON,ENGLAND
[2] CHURCHILL HOSP,OXFORD REG PAIN RELIEF UNIT,OXFORD,OXON,ENGLAND
[3] RADCLIFFE INFIRM,OXFORD EYE HOSP,DEPT OPTOMETRY,OXFORD OX2 6HE,OXON,ENGLAND
关键词
D O I
10.1109/10.238467
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Conventional transmission pulse oximetry is a noninvasive technique for the continuous monitoring of arterial oxygen saturation (SaO(2)) from peripheral vascular beds;such as the finger tip or earlobe. In this paper we propose to exploit the unique transparency of the ocular media to make reflectance pulse oximetry measurements on the retinal fundus. This technique potentially offers significant advantages over conventional pulse oximetry, primarily the ability to monitor cerebral, as opposed to peripheral, oxygen saturation. We have developed an in vitro system to simulate the retinal circulation and ocular optics. This system consists of a flexible cuvette located in a model eye and an extracorporeal blood circuit to simulate arterial blood flow. The system was used to investigate the relationship between SaO(2) and the R/IR ratio in reflectance pulse oximetry. To enable in vivo measurements to be made, we also modified a standard haptic contact lens to hold the pulse oximeter probe in front of the pupil. In a preliminary study, the lens was fitted to an awake volunteer and cardiac-synchronous signals were detected by the retinal pulse oximeter.
引用
收藏
页码:817 / 823
页数:7
相关论文
共 20 条
[11]   PULSE OXIMETRY [J].
KELLEHER, JF .
JOURNAL OF CLINICAL MONITORING, 1989, 5 (01) :37-62
[12]  
LAING RA, 1975, INVEST OPHTH VISUAL, V14, P606
[13]   NONINVASIVE PULSE OXIMETRY UTILIZING SKIN REFLECTANCE PHOTOPLETHYSMOGRAPHY [J].
MENDELSON, Y ;
OCHS, BD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (10) :798-805
[14]   AN INVITRO TISSUE MODEL FOR EVALUATING THE EFFECT OF CARBOXYHEMOGLOBIN CONCENTRATION ON PULSE OXIMETRY [J].
MENDELSON, Y ;
KENT, JC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (06) :625-627
[15]  
Neuman M R, 1987, Adv Exp Med Biol, V220, P135
[16]   PULSE OXIMETRY - TECHNICAL ASPECTS OF MACHINE-DESIGN [J].
POLOGE, JA .
INTERNATIONAL ANESTHESIOLOGY CLINICS, 1987, 25 (03) :137-153
[17]   PULSE OXIMETRY - ANALYSIS OF THEORY, TECHNOLOGY, AND PRACTICE [J].
WUKITSCH, MW ;
PETTERSON, MT ;
TOBLER, DR ;
POLOGE, JA .
JOURNAL OF CLINICAL MONITORING, 1988, 4 (04) :290-301
[18]   EVALUATION OF PULSE OXIMETRY [J].
YELDERMAN, M ;
NEW, W .
ANESTHESIOLOGY, 1983, 59 (04) :349-352
[19]   SPECTROPHOTOMETRIC MONITORING OF ARTERIAL OXYGEN-SATURATION IN THE FINGERTIP [J].
YOSHIYA, I ;
SHIMADA, Y ;
TANAKA, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (01) :27-32
[20]  
ZEIMER RC, 1988, ARCH OPHTHALMOL-CHIC, V106, P126