The factors governing the choice of preamplifier type for the vibrating capacitive probe used in contact potential measurements are examined. Two types are compared: a high input impedance voltage amplifier and a current amplifier. The latter has been increasingly used in recent years due to its great advantages in dealing with parasitic input capacitance. We extend previous analyses, elucidating other advantages of the current amplifier. Particularly important are (i) the reduction of spurious microphonic signals, implying lower systematic error, and (ii) the white noise spectrum of its equivalent contact potential noise, which allows random error to be effectively reduced by increased averaging periods.