QUANTITATIVE VELOCITY IMAGES FROM THICK SLAB 2D-PHASE CONTRAST

被引:6
作者
WEBER, DM
WANG, Y
KOROSEC, FR
MISTRETTA, CA
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
[2] UNIV WISCONSIN,DEPT RADIOL,MADISON,WI 53706
关键词
PHASE CONTRAST; VELOCITY MEASUREMENT;
D O I
10.1002/mrm.1910290210
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Thick slab two-dimensional phase contrast (2D PC) angiography provides a large amount of anatomical information in a short acquisition time. Quantitative velocity information is, however, destroyed by the necessary projection dephasing gradient. We present a 2D PC acquisition scheme which retains quantitative velocity information in a thick slab acquisition. A thick slab acquisition produces an image which is a projection through an entire vessel. Accurate velocity measurements must take into account the intravoxel phase cancellation caused by a distribution of velocities within the vessel. In addition to the description of the data acquisition scheme, we outline a method for determining the mean velocity within a vessel having a laminar flow distribution. It is shown that without the consideration for intravoxel phase cancellation, the mean velocity measurement from thick slab images overestimates the true mean velocity by 10%.
引用
收藏
页码:216 / 225
页数:10
相关论文
共 12 条
[1]  
BERNSTEIN MA, 1991, J MAGN RESON IMAGING
[2]   PROJECTION ANGIOGRAMS OF BLOOD LABELED BY ADIABATIC FAST PASSAGE [J].
DIXON, WT ;
DU, LN ;
FAUL, DD ;
GADO, M ;
ROSSNICK, S .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (03) :454-462
[3]   MAGNETIC-RESONANCE ANGIOGRAPHY [J].
DUMOULIN, CL ;
HART, HR .
RADIOLOGY, 1986, 161 (03) :717-720
[4]   3-DIMENSIONAL PHASE-CONTRAST ANGIOGRAPHY [J].
DUMOULIN, CL ;
SOUZA, SP ;
WALKER, MF ;
WAGLE, W .
MAGNETIC RESONANCE IN MEDICINE, 1989, 9 (01) :139-149
[5]   PHASE-CONTRAST MR ANGIOGRAPHY WITH REDUCED ACQUISITION TIME - NEW CONCEPTS IN SEQUENCE DESIGN [J].
HAUSMANN, R ;
LEWIN, JS ;
LAUB, G .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (04) :415-422
[6]  
IKEZAKI Y, 1990, JPN J MR MED, V10, P346
[7]   ECG-OPTIMIZED PHASE-CONTRAST LINE-SCANNED MR ANGIOGRAPHY [J].
KOROSEC, FR ;
MISTRETTA, CA ;
TURSKI, PA .
MAGNETIC RESONANCE IN MEDICINE, 1992, 24 (02) :221-235
[8]   INPLANE VASCULAR IMAGING - PULSE SEQUENCE DESIGN AND STRATEGY [J].
LENZ, GW ;
HAACKE, EM ;
MASARYK, TJ ;
LAUB, G .
RADIOLOGY, 1988, 166 (03) :875-882
[9]  
PELC NJ, 1991, J MAGN RESON IMAGING, V1, P404
[10]  
SPRITZER CE, 1990, RADIOLOGY, V175