EFFECT OF TISSUE HETEROGENEITY ON QUANTIFICATION IN POSITRON EMISSION TOMOGRAPHY

被引:23
作者
BLOMQVIST, G
LAMMERTSMA, AA
MAZOYER, B
WIENHARD, K
机构
[1] KAROLINSKA HOSP, DEPT CLIN NEUROSCI, EXPTL ALCOHOL & DRUG ADDICT RES SECT, S-17176 STOCKHOLM, SWEDEN
[2] HAMMERSMITH HOSP, ROYAL POSTGRAD MED SCH,MRC,CTR CLIN SCI, CYCLOTRON UNIT,PET METHODOL GRP, LONDON, ENGLAND
[3] HOP ROBERT DEBRE, ANTENNE INFORMAT MED, PARIS, FRANCE
[4] HOP ORSAY, DEPT BIOL, CEA, SERV HOSP FREDERIC JOLIOT, PARIS, FRANCE
[5] MAX PLANCK INST NEUROL RES, COLOGNE, GERMANY
来源
EUROPEAN JOURNAL OF NUCLEAR MEDICINE | 1995年 / 22卷 / 07期
关键词
POSITRON EMISSION TOMOGRAPHY; TISSUE HETEROGENEITY; TRACER KINETICS;
D O I
10.1007/BF01254567
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
As a result of the limited spatial resolution of positron emission tomographic scanners, the measurements of physiological parameters are compromised by tissue heterogeneity, The effect of tissue heterogeneity on a number of parameters was studied by simulation and an analytical method. Five common tracer models were assessed. The input and tissue response functions were assumed to be free from noise and systematic errors. The kinetic model was assumed to be perfect. Two components with different kinetics were mixed in different proportions and contrast with respect to the model parameters. Different experimental protocols were investigated. Of three methods investigated for the measurement of cerebral blood flow (CBF) (steady state, dynamic, integral), the second one was least sensitive to errors caused by tissue heterogeneity and the main effect was an underestimation of the distribution volume. With the steady state method, errors in oxygen extraction fraction caused by tissue heterogeneity were always found to be less than the corresponding errors in CBE For myocardial blood flow the steady state method was found to perform better than the bolus method. The net accumulation of substrate (i.e. rCMR(glc) in the case of glucose analogs) was found to be comparatively insensitive to tissue heterogeneity. Individual rate constants such as k(2) and k(3) for efflux and metabolism of the substrate in the pool of unmetabolized substrate in the tissue, respectively, were found to be more sensitive. In studies of radioligand binding, using only tracer doses, the effect of tissue heterogeneity on the parameter k(on). B-max could be considerable. In studies of radioligand binding using a protocol with two experiments, one with high and one with low specific activity, B-max was found to be insensitive while K-d was very sensitive to tissue heterogeneity.
引用
收藏
页码:652 / 663
页数:12
相关论文
共 40 条
[1]   NONINVASIVE QUANTIFICATION OF REGIONAL MYOCARDIAL BLOOD-FLOW IN CORONARY-ARTERY DISEASE WITH OXYGEN-15-LABELED CARBON-DIOXIDE INHALATION AND POSITRON EMISSION TOMOGRAPHY [J].
ARAUJO, LI ;
LAMMERTSMA, AA ;
RHODES, CG ;
MCFALLS, EO ;
IIDA, H ;
RECHAVIA, E ;
GALASSI, A ;
DESILVA, R ;
JONES, T ;
MASERI, A .
CIRCULATION, 1991, 83 (03) :875-885
[2]  
ARNETT CD, 1986, J NUCL MED, V27, P1878
[3]   MODELS FOR INVIVO KINETIC INTERACTIONS OF DOPAMINE D2-NEURORECEPTORS AND 3-(2'-[F-18]FLUOROETHYL)SPIPERONE EXAMINED WITH POSITRON EMISSION TOMOGRAPHY [J].
BAHN, MM ;
HUANG, SC ;
HAWKINS, RA ;
SATYAMURTHY, N ;
HOFFMAN, JM ;
BARRIO, JR ;
MAZZIOTTA, JC ;
PHELPS, ME .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1989, 9 (06) :840-849
[4]   3-(2'-[F-18]FLUOROETHYL)SPIPERONE - INVIVO BIOCHEMICAL AND KINETIC CHARACTERIZATION IN RODENTS, NONHUMAN-PRIMATES, AND HUMANS [J].
BARRIO, JR ;
SATYAMURTHY, N ;
HUANG, SC ;
KEEN, RE ;
NISSENSON, CHK ;
HOFFMAN, JM ;
ACKERMANN, RF ;
BAHN, MM ;
MAZZIOTTA, JC ;
PHELPS, ME .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1989, 9 (06) :830-839
[5]  
BECK JV, PARAMETER ESTIMATION, pCH7
[6]   NONINVASIVE QUANTITATION OF MYOCARDIAL BLOOD-FLOW IN HUMAN-SUBJECTS WITH OXYGEN-15-LABELED WATER AND POSITRON EMISSION TOMOGRAPHY [J].
BERGMANN, SR ;
HERRERO, P ;
MARKHAM, J ;
WEINHEIMER, CJ ;
WALSH, MN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :639-652
[7]   MAPS OF RECEPTOR-BINDING PARAMETERS IN THE HUMAN BRAIN - A KINETIC-ANALYSIS OF PET MEASUREMENTS [J].
BLOMQVIST, G ;
PAULI, S ;
FARDE, L ;
ERIKSSON, L ;
PERSSON, A ;
HALLDIN, C .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1990, 16 (4-6) :257-265
[8]   POSITRON EMISSION TOMOGRAPHIC MEASUREMENTS OF CEREBRAL GLUCOSE-UTILIZATION USING [1-C-11]D-GLUCOSE [J].
BLOMQVIST, G ;
STONEELANDER, S ;
HALLDIN, C ;
ROLAND, PE ;
WIDEN, L ;
LINDQVIST, M ;
SWAHN, CG ;
LANGSTROM, B ;
WIESEL, FA .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1990, 10 (04) :467-483
[9]  
BUSTANY P, 1981, J CEREB BLOOD FLO S1, V1, pS17
[10]   SIMULATION OF COMPARTMENTAL-MODELS FOR KINETIC DATA FROM A POSITRON EMISSION TOMOGRAPH [J].
COXSON, PG ;
SALMERON, EM ;
HUESMAN, RH ;
MAZOYER, BM .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1992, 37 (03) :205-214