AN ADJUSTABLE RF COIL LOADING DEVICE

被引:8
作者
HAYES, CE
机构
[1] Applied Science Laboratory, General Electric Medical Systems, Waukesha
关键词
RF COIL LOADER; PATIENT LOSS SIMULATOR;
D O I
10.1016/0730-725X(93)90415-A
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper describes an adjustable loading device that can substitute for the tissue losses of various sized patients in a whole-body MR imager. It resembles a lowpass birdcage resonator but with resistors replacing the capacitors on the cylindrical surface. Power dissipated is a monotonic function of the total surface conductance of the loader. The loader can be used in conjunction with a low loss water filled phantom to measure signal-to-noise ratios (SNR). The impact on measured SNR due to screening by the loader and changing the phantom size or composition are also briefly discussed.
引用
收藏
页码:81 / 86
页数:6
相关论文
共 13 条
  • [11] Edelstein, Bottomley, Pfeifer, A signal-to-noise calibration procedure for NMR imaging systems, Med. Phys., 11, pp. 180-185, (1984)
  • [12] Jennings, Lerski, McRobbie, Straughan, Orr, Simulation of patient loading in nuclear resonance imaging through the use of inductive damping loops, Br. J. Radiol., 59, pp. 1031-1034, (1986)
  • [13] Hayes, Edelstein, Schenck, Mueller, Eash, An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5T, J. Magn. Reson., 63, pp. 622-628, (1985)