COMPARISON OF THE KEELER PULSAIR-2000 NONCONTACT TONOMETER WITH GOLDMANN APPLANATION

被引:24
作者
MOSELEY, MJ
THOMPSON, JR
DEUTSCH, J
MISSON, GP
NAYLOR, G
TANYEE, A
TAYLOR, RH
FIELDER, AR
机构
[1] UNIV BIRMINGHAM,DEPT OPHTHALMOL,BIRMINGHAM B15 2TT,W MIDLANDS,ENGLAND
[2] UNIV LEICESTER,LEICESTER ROYAL INFIRM,DEPT OPHTHALMOL,LEICESTER LE1 7RH,ENGLAND
关键词
GOLDMANN APPLANATION TONOMETRY; INTRAOCULAR PRESSURE; NONCONTACT TONOMETRY; PULSAIR-2000; TONOMETRY;
D O I
10.1038/eye.1993.26
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The Pulsair 2000 non-contact tonometer (Keeler Ltd, UK) is compared with the Goldmann applanation tonometer. Data from 80 eyes were acquired by four experienced observers. A linear regression analysis showed the relationship between the instruments to be: Pulsair=0.66 + 0.95 Goldmann. Individual components of variation were analysed by analysis of variance which indicated a significant variation in the slope of the regression equation due to observers (p=0.02) but not to the order in which topical anaesthesia was administered. Differences between two Pulsair instruments were of marginal significance (p=0.07). The intercept of the regression equation was unaffected by any of the components of variation. Seventy-nine per cent of averaged intraocular pressure measurements obtained with the Pulsair 2000 fell on or within +/-3 mmHg of those measured with the Goldmann tonometer. It is concluded that the Pulsair 2000 can provide clinically useful measurements of intraocular pressure.
引用
收藏
页码:127 / 130
页数:4
相关论文
共 36 条
[11]   A NEW HAND-HELD AIR IMPULSE TONOMETER [J].
FISHER, JH ;
WATSON, PG ;
SPAETH, G .
EYE, 1988, 2 :238-242
[12]  
GOLDMANN H, 1958, AM J OPHTHALMOL, V48, P213
[13]  
GROLMAN B, 1972, AMER J OPT ARCH AM A, V49, P646
[14]  
HAAGSTREIT AG, OPHTHALMOLOGICAL INS
[15]  
KAO SF, 1987, OPHTHALMOLOGY, V94, P1541
[16]  
KOHL P, 1989, Investigative Ophthalmology and Visual Science, V30, P241
[17]  
KRAKAU CET, 1971, ACTA OPHTHALMOL, V49, P611
[18]  
LAGERLOF O, 1990, ACTA OPHTHALMOL, V68, P221
[19]   A NEW NONCONTACT TONOMETER [J].
LANGMANN, G ;
SCHUHMANN, G ;
LANGMANN, A ;
ZENZ, H .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1990, 197 (05) :393-396
[20]   COMPARISON OF A NEW NON-CONTACT TONOMETER WITH GOLDMANN APPLANATION [J].
MOSELEY, MJ ;
EVANS, NM ;
FIELDER, AR .
EYE, 1989, 3 :332-337