SILICONE IMPLANT RUPTURE DIAGNOSIS USING COMPUTED-TOMOGRAPHY - A CASE-REPORT AND EXPERIENCE WITH 22 SURGICALLY REMOVED IMPLANTS

被引:10
作者
AHN, CY
DEBRUHL, ND
GORCZYCA, DP
BASSETT, LW
SHAW, WW
机构
[1] UNIV CALIF LOS ANGELES,MED CTR,DIV PLAST & RECONSTRUCT SURG,LOS ANGELES,CA 90024
[2] UNIV CALIF LOS ANGELES,MED CTR,DEPT RADIOL SCI,LOS ANGELES,CA 90024
关键词
D O I
10.1097/00000637-199412000-00011
中图分类号
R61 [外科手术学];
学科分类号
摘要
Silicone elastomer shell rupture is a complication of silicone implants. To date, the rate of implant rupture has not been well documented. Magnetic resonance imaging and sonography are noninvasive breast implant imaging modalities that have been shown to be useful in evaluating the integrity of implants. We present a case of rupture detection using a follow-up computed tomographic (CT) scan of a breast cancer patient, which prompted us to use CT scans to evaluate explants of patients undergoing implant removal surgery. The purpose of the investigation was to evaluate the effectiveness of CT scan in detecting rupture. CT scan was performed on 22 explants with intact capsules, for which 17 ruptures were confirmed: 16 true-positive ruptures, 5 true-negative ruptures, 0 false-positive ruptures, and 1 false-negative rupture were identified. CT scan was shown to be highly sensitive and specific in rupture detection, comparable to magnetic resonance imaging. Although CT scans are consistently reliable, patients are exposed to ionizing radiation; therefore, it is not recommended for patients with augmentation mammoplasty. This study characterizes the appearance of implant rupture on CT scan, which may be useful in evaluating breast cancer patients reconstructed with silicone implants.
引用
收藏
页码:624 / 628
页数:5
相关论文
共 16 条
[1]   REGIONAL SILICONE-GEL MIGRATION IN PATIENTS WITH RUPTURED IMPLANTS [J].
AHN, CY ;
SHAW, WW .
ANNALS OF PLASTIC SURGERY, 1994, 33 (02) :201-208
[2]   DEFINITIVE DIAGNOSIS OF BREAST IMPLANT RUPTURE USING MAGNETIC-RESONANCE-IMAGING [J].
AHN, CY ;
SHAW, WW ;
NARAYANAN, K ;
GORCZYCA, DP ;
SINHA, S ;
DEBRUHL, ND ;
BASSETT, LW .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1993, 92 (04) :681-691
[3]  
AHN CY, 1994, PLAST RECONSTR SURG, V94, P1
[4]   BLEEDING OF SILICONE FROM BAG-GEL BREAST IMPLANTS, AND ITS CLINICAL RELATION TO FIBROUS CAPSULE REACTION [J].
BARKER, DE ;
RETSKY, MI ;
SCHULTZ, S .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1978, 61 (06) :836-841
[5]   HUMAN ADJUVANT DISEASE FOLLOWING AUGMENTATION MAMMOPLASTY [J].
BROZENA, SJ ;
FENSKE, NA ;
CRUSE, CW ;
ESPINOZA, CG ;
VASEY, FB ;
GERMAIN, BF ;
ESPINOZA, LR .
ARCHIVES OF DERMATOLOGY, 1988, 124 (09) :1383-1386
[6]   SILICONE BREAST IMPLANTS - US EVALUATION [J].
DEBRUHL, ND ;
GORCZYCA, DP ;
AHN, CY ;
SHAW, WW ;
BASSETT, LW .
RADIOLOGY, 1993, 189 (01) :95-98
[7]   AUGMENTATION MAMMAPALSTY - NORMAL AND ABNORMAL FINDINGS WITH MAMMOGRAPHY AND ULTRASOUND [J].
GANOTT, MA ;
HARRIS, KM ;
ILKHANIPOUR, ZS ;
COSTAGRECO, MA .
RADIOGRAPHICS, 1992, 12 (02) :281-295
[8]   SILICONE BREAST IMPLANT RUPTURES IN AN ANIMAL-MODEL - COMPARISON OF MAMMOGRAPHY, MR-IMAGING, US, AND CT [J].
GORCZYCA, DP ;
DEBRUHL, ND ;
AHN, CY ;
HOYT, A ;
SAYRE, JW ;
NUDELL, PP ;
MCCOMBS, M ;
SHAW, WW ;
BASSETT, LW .
RADIOLOGY, 1994, 190 (01) :227-232
[9]   CLINICAL SPECTRUM OF CONNECTIVE-TISSUE DISEASE AFTER COSMETIC SURGERY - OBSERVATIONS ON 18 PATIENTS AND A REVIEW OF THE JAPANESE LITERATURE [J].
KUMAGAI, Y ;
SHIOKAWA, Y ;
MEDSGER, TA ;
RODNAN, GP .
ARTHRITIS AND RHEUMATISM, 1984, 27 (01) :1-12
[10]   LOCALIZED MORPHEA AFTER SILICONE GEL BREAST IMPLANTATION - MORE EVIDENCE FOR A CAUSE-AND-EFFECT RELATIONSHIP [J].
LAZAR, AP ;
LAZAR, P .
ARCHIVES OF DERMATOLOGY, 1991, 127 (02) :263-263