MEASUREMENTS OF THE ABSOLUTE SPECTRAL SENSITIVITY OF X-RAY SEMICONDUCTOR-DETECTORS IN THE PHOTON ENERGY-RANGE OF 1.5-15 KEV USING WHITE SR BEAM OF THE VEPP-3 STORAGE-RING

被引:8
作者
DOLBNYA, IP [1 ]
MAKAROV, OA [1 ]
MEZENTSEV, NA [1 ]
PINDYURIN, VF [1 ]
SUBBOTIN, AN [1 ]
机构
[1] INST EXPTL PHYS,ARZAMAS 607200,RUSSIA
关键词
D O I
10.1016/0168-9002(94)01672-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Results of measurements of the absolute spectral sensitivity of silicon semiconductor detectors in the X-ray quanta energy range of 1.5-15 keV are presented. The detectors, being calibrated, were placed into the direct ''white'' synchrotron radiation (SR) beam from the VEPP-3 storage ring. The spectrum of X-radiation at the entrance window of the detectors was changed by using sets of calibrated filters, as well as by varying the energy of the electrons in the storage ring. The possibility of accurate calculation of the SR spectrum on the calibrated detector under its irradiation in different conditions allowed us to determine the detector spectral sensitivity from a set of integral equations connecting the spectral sensitivity to the registered detector currents. The analysis of possible experimental errors indicates that the absolute spectral sensitivity of the detectors was restored with an accuracy of not worse than 10% in the total photon energy range under the study.
引用
收藏
页码:427 / 431
页数:5
相关论文
共 10 条
[1]   ELECTRON STORAGE RING BESSY AS A SOURCE OF CALCULABLE SPECTRAL PHOTON FLUX IN THE X-RAY REGION [J].
ARNOLD, D ;
ULM, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :1539-1542
[2]   X-RAY-LITHOGRAPHY AT THE VEPP-3 STORAGE RING [J].
ARTAMONOVA, LD ;
GENTSELEV, AN ;
DEIS, GA ;
KRASNOPEROVA, AA ;
MIKHALYOV, EV ;
PROKOPENKO, VS ;
KULIPANOV, GN ;
MEZENTSEVA, LA ;
PINDYURIN, VF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :764-766
[3]   A NEW METHOD OF CONSTRAINED OPTIMIZATION AND A COMPARISON WITH OTHER METHODS [J].
BOX, MJ .
COMPUTER JOURNAL, 1965, 8 (01) :42-52
[4]   MEASUREMENTS OF ABSOLUTE SENSITIVITY OF SEMICONDUCTOR-DETECTORS FOR 7-20 KEV X-RAYS USING SYNCHROTRON RADIATION OF VEPP-3 STORAGE RING [J].
DOLBNYA, IP ;
MEZENTSEV, NA ;
PINDYURIN, VF ;
ROMANENKO, KP ;
SUBBOTIN, AN .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :685-688
[5]   RESPONSE OF AMORPHOUS-SILICON P-I-N DETECTORS TO IONIZING PARTICLES [J].
DUBEAU, J ;
POCHET, T ;
HAMEL, LA ;
EQUER, B ;
KARAR, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1991, 54 (04) :458-471
[6]  
GREEN GK, 1977, BNL50595 REPR
[7]  
KUNZ C, 1979, SYNCHROTRON RAD TECH
[8]  
SKRINSKY AN, 1977, USP FIZ NAUK, V122, P369
[9]  
VERETENNIKOV AI, 1985, METHODS PULSE RAD RE
[10]  
1994, REPORT ACTIVITIES SI, P55