PREDICTIVE RESPIRATORY GATING - A NEW METHOD TO REDUCE MOTION ARTIFACTS ON CT SCANS

被引:106
作者
RITCHIE, CJ
HSIEH, J
GARD, MF
GODWIN, JD
KIM, YM
CRAWFORD, CR
机构
[1] UNIV WASHINGTON,DEPT RADIOL,SEATTLE,WA 98195
[2] UNIV WASHINGTON,DEPT ELECT ENGN,SEATTLE,WA 98195
[3] GE CO,MED SYST,MILWAUKEE,WI 53201
关键词
COMPUTED TOMOGRAPHY (CT); ARTIFACT; PHYSICS; TECHNOLOGY;
D O I
10.1148/radiology.190.3.8115638
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
PURPOSE: To evaluate a gating system, called predictive respiratory gating (PRG), that reduces motion-induced artifacts on computed tomographic (CT) scans of patients who cannot suspend respiration. MATERIALS AND METHODS: PRG uses a respiration monitor and a new algorithm to predict when a motionless period is about to occur. It automatically starts scanning so the scan is temporally centered around the motionless period at end inspiration or end expiration. To demonstrate PRG, CT was performed on a motion phantom and a quietly breathing volunteer with and without gating. RESULTS: Scans of the phantom obtained with PRG contained less motion-induced streaking and blurring than did scans acquired without PRG. Scans of the volunteer gated at end expiration contained significantly less artifact than nongated scans (P < .03). CONCLUSION: PRG reduced motion artifact on scans of a spontaneously breathing volunteer. PRG may be able to reduce motion artifacts on scans of patients unable to suspend respiration.
引用
收藏
页码:847 / 852
页数:6
相关论文
共 15 条
[1]
THE RF COIL AS A SENSITIVE MOTION DETECTOR FOR MAGNETIC-RESONANCE IMAGING [J].
BUIKMAN, D ;
HELZEL, T ;
ROSCHMANN, P .
MAGNETIC RESONANCE IMAGING, 1988, 6 (03) :281-289
[2]
REDUCTION OF MOTION ARTIFACTS IN COMPUTED-TOMOGRAPHY [J].
CRAWFORD, CR ;
GODWIN, JD ;
PELC, NJ .
IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 :485-486
[3]
SPIRAL CT OF THE PANCREAS [J].
DUPUY, DE ;
COSTELLO, P ;
ECKER, CP .
RADIOLOGY, 1992, 183 (03) :815-818
[4]
GOULD RG, 1992, ULTRAFAST COMPUTED T, P1
[5]
SEMIAUTOMATIC EVALUATION PROCEDURES FOR QUANTITATIVE CT OF THE LUNG [J].
KALENDER, WA ;
FICHTE, H ;
BAUTZ, W ;
SKALEJ, M .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (02) :248-255
[6]
THE DOUBLE-FISSURE SIGN - A MOTION ARTIFACT ON THIN-SECTION CT SCANS [J].
MAYO, JR ;
MULLER, NL ;
HENKELMAN, RM .
RADIOLOGY, 1987, 165 (02) :580-581
[7]
PROSPECTIVELY GATED CARDIAC COMPUTED-TOMOGRAPHY [J].
MOORE, SC ;
JUDY, PF ;
GARNIC, JD ;
KAMBIC, GX ;
BONK, F ;
COCHRAN, G ;
MARGOSIAN, P ;
MCCROSKEY, W ;
FOOTE, F .
MEDICAL PHYSICS, 1983, 10 (06) :846-855
[8]
LOW-DOSE CT OF THE LUNGS - PRELIMINARY-OBSERVATIONS [J].
NAIDICH, DP ;
MARSHALL, CH ;
GRIBBIN, C ;
ARAMS, RS ;
MCCAULEY, DI .
RADIOLOGY, 1990, 175 (03) :729-731
[9]
RITCHIE CJ, 1992, P IEEE EMBS, V14, P1782
[10]
MINIMUM SCAN SPEEDS FOR SUPPRESSION OF MOTION ARTIFACTS IN CT [J].
RITCHIE, CJ ;
GODWIN, JD ;
CRAWFORD, CR ;
STANFORD, W ;
ANNO, H ;
KIM, Y .
RADIOLOGY, 1992, 185 (01) :37-42