ANALYSIS OF SILICATE ROCKS USING FIELD-PORTABLE X-RAY-FLUORESCENCE INSTRUMENTATION INCORPORATING A MERCURY(II) IODIDE DETECTOR - A PRELIMINARY ASSESSMENT OF ANALYTICAL PERFORMANCE

被引:59
作者
POTTS, PJ [1 ]
WEBB, PC [1 ]
WILLIAMSTHORPE, O [1 ]
KILWORTH, R [1 ]
机构
[1] THERMO ELECTRON LTD,DIV ANALYT INSTRUMENTS,WARRINGTON WA3 7QN,CHESHIRE,ENGLAND
关键词
PORTABLE ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS SYSTEM; MERCURY(II) IODIDE DETECTOR; RADIOISOTOPE EXCITATION; ACCURACY; PRECISION; DETECTION LIMIT;
D O I
10.1039/an9952001273
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An assessment is reported of the performance of a field-portable energy dispersive X-ray fluorescence analysis system using Fe-55, Cd-109 and Am-241 excitation sources and a high resolution mercury(II) iodide detector, Seventy international reference materials, mainly of silicate composition, were analysed as compressed powder pellets. Results from spectra recorded for live times of 200 s (per source) showed that the major elements K, Ca, Ti, Mn and Fe and the trace elements Ba, Nb, Rb, Sr, Y and Zr could be determined routinely in 'normal' silicate rocks and in addition Co, Cr, Cu, Ga, La, Nd, Ni, V and Zn at higher concentrations or in mineralized samples, Detection limits for the most sensitively measured trace elements (Rb, Sr, Y, Zr, Nb) were found to be in the range 6-14 mu g g(-1) and for Ba (K-line), 21 mu g g(-1) Precision in the determination of the major elements was generally in the range 0.45-2% (relative standard deviation) and a high degree of accuracy was achieved when judged from a self-consistent comparison of analysed values with the expected compositions of the 70 reference materials, Having characterized performance in the laboratory with 'ideal' control over sample presentation, work is now in progress to evaluate additional discrepancies that will be encountered in the analysis of geological and archaeological samples in the field.
引用
收藏
页码:1273 / 1278
页数:6
相关论文
共 24 条
[11]   MONS-CLAUDIANUS AND THE PROBLEM OF THE GRANITO-DEL-FORO - A GEOLOGICAL AND GEOCHEMICAL APPROACH [J].
PEACOCK, DPS ;
WILLIAMSTHORPE, O ;
THORPE, RS ;
TINDLE, AG .
ANTIQUITY, 1994, 68 (259) :209-230
[12]  
POTTS P. J., 1992, GEOCHEMICAL REFERENC
[13]   CALCULATION OF CALIBRATION LINE PARAMETERS FROM REFERENCE MATERIAL DATA IN THE ANALYSIS OF SILICATE ROCKS - THEIL INCOMPLETE METHOD COMPARED WITH LEAST-SQUARES REGRESSION [J].
POTTS, PJ ;
WRIGHT, DW ;
WATSON, JS ;
WEBB, PC .
CHEMICAL GEOLOGY, 1987, 63 (3-4) :345-354
[14]   X-RAY-FLUORESCENCE SPECTROMETRY [J].
POTTS, PJ ;
WEBB, PC .
JOURNAL OF GEOCHEMICAL EXPLORATION, 1992, 44 (1-3) :251-296
[15]   ENERGY-DISPERSIVE X-RAY-FLUORESCENCE ANALYSIS OF SILICATE ROCKS - COMPARISONS WITH WAVELENGTH-DISPERSIVE PERFORMANCE [J].
POTTS, PJ ;
WEBB, PC ;
WATSON, JS .
ANALYST, 1985, 110 (05) :507-513
[16]   ENERGY-DISPERSIVE X-RAY-FLUORESCENCE ANALYSIS OF SILICATE ROCKS FOR MAJOR AND TRACE-ELEMENTS [J].
POTTS, PJ ;
WEBB, PC ;
WATSON, JS .
X-RAY SPECTROMETRY, 1984, 13 (01) :2-15
[17]   ZIRCONIUM DETERMINATION BY ED-XRF - A CRITICAL-EVALUATION OF SILICATE REFERENCE MATERIALS AS CALIBRATION STANDARDS [J].
POTTS, PJ ;
WEBB, PC ;
WATSON, JS .
GEOSTANDARDS NEWSLETTER, 1990, 14 (01) :127-136
[18]  
RAAB GA, 1991, ENVIR SCI R, V42, P155
[19]  
SPARKS CJ, 1976, ADV XRAY ANALYSIS, V19, P19
[20]   TRACE-ELEMENT ANALYSIS OF GEOCHEMICAL REFERENCE SAMPLES BY ENERGY DISPERSIVE-X-RAY FLUORESCENCE SPECTROMETRY [J].
WEBB, PC ;
POTTS, PJ ;
WATSON, JS .
GEOSTANDARDS NEWSLETTER, 1990, 14 (03) :361-372