FLEXIBLE (KAPTON-BASED) MICROSENSOR ARRAYS OF HIGH-STABILITY FOR CARDIOVASCULAR APPLICATIONS

被引:91
作者
LINDNER, E
COSOFRET, VV
UFER, S
BUCK, RP
KUSY, RP
ASH, RB
NAGLE, HT
机构
[1] UNIV N CAROLINA CHAPEL HILL, DEPT CHEM, CHAPEL HILL, NC 27599 USA
[2] UNIV N CAROLINA CHAPEL HILL, DEPT ORTHODONT & BIOMED ENGN, CHAPEL HILL, NC 27599 USA
[3] N CAROLINA STATE UNIV, DEPT ELECT & COMP ENGN, RALEIGH, NC 27695 USA
来源
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS | 1993年 / 89卷 / 02期
关键词
D O I
10.1039/ft9938900361
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The design, fabrication and performance characteristics of Kapton-based planar mini and semimicro potentiometric sensors with an Ag/AgCl or a quinhydrone-based redox internal reference electrode are described. The ion-selective membranes cast from conventional and various modified PVC matrices and containing different pH-sensitive ionophores are ranked on the basis of their performances in hostile environments. The adhesive bonding strength of the different PVC membranes to the polyimide-coated Kapton substrate was quantitatively evaluated as a function of fabrication procedure and sample solution contact. The long-term stability of the electrodes was characterized by the alterations of the analytical parameters of the sensors over a period of time, as well as by determining the resistance changes of their sensing membranes.
引用
收藏
页码:361 / 367
页数:7
相关论文
共 29 条
  • [11] Ives D. J., 1961, REFERENCE ELECTRODES
  • [12] Janata J., 1989, PRINCIPLES CHEM SENS
  • [13] FABRICATION, EVALUATION, AND USE OF EXTRACELLULAR K+ AND H+ ION-SELECTIVE ELECTRODES
    JOHNSON, TA
    ENGLE, CL
    KUSY, RP
    KNISLEY, SB
    GRAEBNER, CA
    GETTES, LS
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 258 (04): : H1224 - H1231
  • [14] THIN-FILM ION-SELECTIVE SENSORS BASED ON NEUTRAL CARRIER MEMBRANES
    KEPLINGER, F
    GLATZ, R
    JACHIMOWICZ, A
    URBAN, G
    KOHL, F
    OLCAYTUG, F
    PROHASKA, OJ
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) : 272 - 274
  • [15] KO WH, 1985, IMPLANTABLE SENSORS, P259
  • [16] KUSY RP, UNPUB
  • [17] LINDNER E, UNPUB ANAL CHEM
  • [18] MA SC, 1990, MIKROCHIM ACTA, V1, P197
  • [19] OEHME F, 1986, GIT FACHZ LAB, P595
  • [20] LIFE TIME OF NEUTRAL CARRIER BASED ION-SELECTIVE LIQUID-MEMBRANE ELECTRODES
    OESCH, U
    SIMON, W
    [J]. ANALYTICAL CHEMISTRY, 1980, 52 (04) : 692 - 700