AN ELECTROCHEMICAL CAPILLARY FILL DEVICE FOR THE ANALYSIS OF GLUCOSE INCORPORATING GLUCOSE-OXIDASE AND RUTHENIUM(III) HEXAMINE AS MEDIATOR

被引:83
作者
MORRIS, NA
CARDOSI, MF
BIRCH, BJ
TURNER, APF
机构
[1] CRANFIELD INST TECHNOL,CTR BIOTECHNOL,CRANFIELD MK43 0AL,BEDS,ENGLAND
[2] PAISLEY COLL TECHNOL,DEPT BIOL,PAISLEY PA1 2BE,RENFREW,SCOTLAND
[3] UNILEVER RES,COLWORTH LAB,SHARNBROOK MK44 1LQ,BEDS,ENGLAND
关键词
GLUCOSE; BIOSENSOR; MEDIATOR;
D O I
10.1002/elan.1140040104
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article describes a coulometric measurement device for glucose based on thin-layer bulk electrolysis. The analysis is carried out in a specially constructed cell, the electrochemical capillary fill device, which has a total volume of less than 20-mu-l. Once the sample enters the cell by capillary action, the freeze-dried assay components (glucose oxidase, ruthenium (III) hexamine, buffer salts, etc.) become solubilized and quickly disperse throughout the volume of the cell. After a period of one min, the time taken for the enzyme catalyzed oxidation of glucose to gluconic acid to reach completion, a potential step is applied and the current transient, resulting from the oxidation of reduced mediator at the working electrode, integrated over a period of 60 s. From the resultant charge value, the concentration of glucose in the sample can be unambiguously determined. The response of the sensor is both linear and accurate over the range 2-30 mM, which covers the normally accepted range for blood glucose levels in diabetic patients. The second order rate constant for the reaction between reduced glucose oxidase and ruthenium (III) hexamine was evaluated by conducting a numerically based analysis that was designed to separate the diffusional and catalytic components of current transients.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 22 条
[11]  
Kuwana T, 1977, ACS SYM SER, V38, P143
[12]   ELECTROCHEMICAL STUDY OF SULFONATED FERROCENES AS REDOX MEDIATORS IN ENZYME ELECTRODES [J].
LIAUDET, E ;
BATTAGLINI, F ;
CALVO, EJ .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1990, 293 (1-2) :55-68
[13]  
MATTHEWS DR, 1987, LANCET, V1, P778
[14]   THEORY OF STATIONARY ELECTRODE POLAROGRAPHY - SINGLE SCAN + CYCLIC METHODS APPLIED TO REVERSIBLE IRREVERSIBLE + KINETIC SYSTEMS [J].
NICHOLSON, RS ;
SHAIN, I .
ANALYTICAL CHEMISTRY, 1964, 36 (04) :706-&
[15]   A STUDY OF INTERFERENCES IN GLUCOSE MEASUREMENTS IN BLOOD BY HYDROGEN-PEROXIDE BASED GLUCOSE PROBES [J].
PALLESCHI, G ;
RAHNI, MAN ;
LUBRANO, GJ ;
NGWAINBI, JN ;
GUILBAULT, GG .
ANALYTICAL BIOCHEMISTRY, 1986, 159 (01) :114-121
[16]  
RENNIE IDB, 1964, LANCET, V2, P884
[17]  
SAVEANT JM, 1960, ADV POLAROGR, V1, P10
[18]  
Skyler J S, 1978, Diabetes Care, V1, P150, DOI 10.2337/diacare.1.3.150
[19]  
SONKSON PH, 1978, LANCET, V1, P29
[20]  
STROUD KA, 1981, ENG MATH