THROMBOGENIC PROPERTIES OF UNTREATED AND POLY(ETHYLENE OXIDE)-MODIFIED POLYMERIC MATRICES USEFUL FOR PREPARING INTRAARTERIAL ION-SELECTIVE ELECTRODES

被引:93
作者
ESPADASTORRE, C [1 ]
MEYERHOFF, ME [1 ]
机构
[1] UNIV MICHIGAN,DEPT CHEM,ANN ARBOR,MI 48109
关键词
D O I
10.1021/ac00114a003
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In vitro platelet adhesion studies are used to compare the thrombogenic properties of various polymer matrices useful for preparing implantable ion-selective membrane electrodes. Conventional plasticized poly(vinyl chloride) and alternate polyurethane materials (Tecoflex, Pellethane) doped with proton- (tridodecylamine) and potassium-selective (valinomycin) ionophores are shown to be potentially thrombogenic. Incorporation of high molecular weight block copolymers of poly(ethylene oxide) and poly(propylene oxide) (e.g., Pluronic F108 and Tetronic 1508) within ion-selective membranes reduces platelet adhesion. A more marked decrease in platelet adhesion is, however, observed when the Tecoflex-based membranes are coated with a thin photo-cross-linked layer of poly(ethylene oxide). Such surface-modified membranes are shown to retain potentiometric ion response properties (i.e., selectivity, response times, response slopes, etc.) essentially equivalent to untreated membranes.
引用
收藏
页码:3108 / 3114
页数:7
相关论文
共 42 条
[1]   EXTRACTION OF AUROCYANIDE ION-PAIRS BY POLY(OXYETHYLENE) EXTRACTANTS [J].
ADAMS, MD ;
WADE, PW ;
HANCOCK, RD .
TALANTA, 1990, 37 (09) :875-883
[2]   PREVENTION OF PROTEIN ADSORPTION AND PLATELET-ADHESION ON SURFACES BY PEO PPO PEO TRIBLOCK COPOLYMERS [J].
AMIJI, M ;
PARK, K .
BIOMATERIALS, 1992, 13 (10) :682-692
[3]  
AMMANN D, 1985, ION MEASUREMENTS PHY, P102
[4]  
ANDERSON J M, 1985, Critical Reviews in Biocompatibility, V1, P111
[5]  
BAILEY PL, 1980, ANAL IONSELECTIVE EL
[6]   OPTIMUM COMPOSITION OF NEUTRAL CARRIER BASED PH ELECTRODES [J].
BAKKER, E ;
XU, AP ;
PRETSCH, E .
ANALYTICA CHIMICA ACTA, 1994, 295 (03) :253-262
[7]   PLATELET DEPOSITION STUDIES ON COPOLYETHER URETHANES MODIFIED WITH POLY (ETHYLENE-OXIDE) [J].
BRINKMAN, E ;
POOT, A ;
VANDERDOES, L ;
BANTJES, A .
BIOMATERIALS, 1990, 11 (03) :200-205
[8]   ELECTROCHEMICAL PERFORMANCE, BIOCOMPATIBILITY, AND ADHESION OF NEW POLYMER MATRICES FOR SOLID-STATE ION SENSORS [J].
CHA, GS ;
LIU, D ;
MEYERHOFF, ME ;
CANTOR, HC ;
MIDGLEY, AR ;
GOLDBERG, HD ;
BROWN, RB .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1666-1672
[9]   BIOCOMPATIBLE NEEDLE-TYPE GLUCOSE SENSOR WITH POTENTIAL FOR USE IN-VIVO [J].
CHEN, CY ;
SU, YC ;
ISHIHARA, K ;
NAKABAYASHI, N ;
TAMIYA, E ;
KARUBE, I .
ELECTROANALYSIS, 1993, 5 (04) :269-276
[10]   POTENTIOMETRIC COMBINATION ION CARBON-DIOXIDE SENSORS FOR INVITRO AND INVIVO BLOOD MEASUREMENTS [J].
COLLISON, ME ;
AEBLI, GV ;
PETTY, J ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1989, 61 (21) :2365-2372