Water samples with variable salinity are analysed by a flow-injection-inductively coupled plasma (ICP) system. The sampling zone approach is used to level the sample and standard saline contents in order to compensate for matrix effects. Calibration graphs for Ca, Mg, Al, Fe, P and Si at three different sodium concentrations are obtained. The sample and standard salinity matching is done automatically by processing the sample sodium signal from the ICP with a microcomputer and adding a suitable amount of sodium. The decision process consists in choosing the appropriate standard solution and adjusting the commutation delay for the sampling zone, using a previously determined time-concentration curve. Sodium interference is minimized, allowing the analysis of water samples with a saline matrix up to 2.5% (w/v) NaCl. The proposed method is suitable for 60 samples h-1. © 1990.
引用
收藏
页码:207 / 212
页数:6
相关论文
共 14 条
[1]   SPECTROMETRIC STUDY OF HIGH-FREQUENCY INDUCED PLASMA .2. DIFFERENT TYPES OF INTERFERENCES OBSERVED [J].
ABDALLAH, MH ;
MERMET, JM ;
TRASSY, C .
ANALYTICA CHIMICA ACTA, 1976, 87 (02) :329-339
[2]   SIMULTANEOUS DETERMINATION OF 5 TRACE-METALS IN SEAWATER BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY WITH ULTRASONIC NEBULIZATION [J].
BERMAN, SS ;
MCLAREN, JW ;
WILLIE, SN .
ANALYTICAL CHEMISTRY, 1980, 52 (03) :488-492
[3]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[4]   MULTI-ELEMENT SIMPLEX OPTIMIZATION FOR INDUCTIVELY-COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY WITH A PLASMA TORCH FITTED WITH A WIDE-BORE INJECTOR TUBE .2. ESTABLISHMENT OF OPTIMUM CONDITIONS FOR MINIMUM INTERFERENCE [J].
EBDON, L ;
CARPENTER, R .
ANALYTICA CHIMICA ACTA, 1988, 209 (1-2) :135-145
[5]   FLOW-INJECTION CALIBRATION OF INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY USING THE GENERALIZED STANDARD ADDITIONS METHOD [J].
GINE, MF ;
KRUG, FJ ;
FILHO, HB ;
DOSREIS, BF ;
ZAGATTO, EAG ;
BRUNS, RE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :673-678
[6]  
GINE MF, 1983, ANAL CHIM ACTA, V115, P131
[7]   SOME ASPECTS OF MATRIX INTERFERENCE CAUSED BY ELEMENTS OF LOW IONIZATION-POTENTIAL IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
GUNTER, WH ;
VISSER, K ;
ZEEMAN, PB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (07) :571-581
[8]   SPATIAL-DISTRIBUTION OF TEMPERATURE AND NUMBER DENSITIES OF ELECTRONS AND ATOMIC AND IONIC SPECIES IN AN INDUCTIVELY COUPLED RF ARGON PLASMA [J].
KORNBLUM, GR ;
DEGALAN, L .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1977, 32 (02) :71-96
[9]   INDUCTIVELY COUPLED PLASMA-OPTICAL EMISSION ANALYTICAL SPECTROMETRY - STUDY OF SOME INTERELEMENT EFFECTS [J].
LARSON, GF ;
FASSEL, VA ;
SCOTT, RH ;
KNISELEY, RN .
ANALYTICAL CHEMISTRY, 1975, 47 (02) :238-243
[10]  
MERMET JM, 1975, ANAL CHIM ACTA, V70, P271