SUBCUTANEOUS GLUCOSE-CONCENTRATION IN HUMANS - REAL ESTIMATION AND CONTINUOUS MONITORING

被引:50
作者
STERNBERG, F
MEYERHOFF, C
MENNEL, FJ
BISCHOF, F
PFEIFFER, EF
机构
[1] Institute for Diabetes-Technology, University of Ulm, Donau
[2] Inst. für Diabetes-Technologie, Universität Ulm, D-89081 Ulm/Donau
关键词
D O I
10.2337/diacare.18.9.1266
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE - To determine the real subcutaneous glucose concentration in healthy volunteers to help in the development of new calibration methods for subcutaneous glucosensors. RESEARCH DESIGN AND METHODS - We developed a new method to estimate the real subcutaneous glucose concentration based on the recirculation of phosphate-buffered saline (PBS) in a microdialysis probe inserted into the subcutaneous tissue. Tissue glucose diffuses into the probe until complete equilibration between the glucose concentration outside and inside the microdialysis probe is achieved. Later, the glucose content of the recirculated PBS is assessed in vitro. We applied the method in 10 healthy volunteers under fasting state and during a hyperglycemic clamp. in addition, we monitored the subcutaneous glucose with an enzymatic-amperometric glucosensor combined with a microdialysis probe. RESULTS - The subcutaneous glucose concentration measured by the recirculation method was 72 +/- 6 and 78 +/- 6% of the blood glucose measured in the fasting state and during the hyperglycemic clamps, respectively. On the other hand, the glucosensor's signal correlated significantly with the blood glucose. CONCLUSION The recirculation method estimated the real subcutaneous glucose concentration, opening the way to develop new calibration procedures for subcutaneous glucosensors. However, a suitable calibration procedure is still lacking.
引用
收藏
页码:1266 / 1269
页数:4
相关论文
共 9 条
  • [1] BRAIN MICRODIALYSIS
    BENVENISTE, H
    [J]. JOURNAL OF NEUROCHEMISTRY, 1989, 52 (06) : 1667 - 1679
  • [2] INVIVO MEASUREMENT OF SUBCUTANEOUS GLUCOSE-CONCENTRATIONS WITH AN ENZYMATIC GLUCOSE SENSOR AND A WICK METHOD
    BRUCKEL, J
    KERNER, W
    ZIER, H
    STEINBACH, G
    PFEIFFER, EF
    [J]. KLINISCHE WOCHENSCHRIFT, 1989, 67 (09): : 491 - 495
  • [3] DEFRONZO RA, 1979, AM J PHYSIOL, V237, pE214
  • [4] ASSESSMENT OF SUBCUTANEOUS GLUCOSE-CONCENTRATION - VALIDATION OF THE WICK TECHNIQUE AS A REFERENCE FOR IMPLANTED ELECTROCHEMICAL SENSORS IN NORMAL AND DIABETIC DOGS
    FISCHER, U
    ERTLE, R
    ABEL, P
    REBRIN, K
    BRUNSTEIN, E
    VONDORSCHE, HH
    FREYSE, EJ
    [J]. DIABETOLOGIA, 1987, 30 (12) : 940 - 945
  • [5] ON LINE CONTINUOUS MONITORING OF SUBCUTANEOUS TISSUE GLUCOSE IN MEN BY COMBINING PORTABLE GLUCOSENSOR WITH MICRODIALYSIS
    MEYERHOFF, C
    BISCHOF, F
    STERNBERG, F
    ZIER, H
    PFEIFFER, EF
    [J]. DIABETOLOGIA, 1992, 35 (11) : 1087 - 1092
  • [6] ON LINE CONTINUOUS MONITORING OF SUBCUTANEOUS TISSUE GLUCOSE IS FEASIBLE BY COMBINING PORTABLE GLUCOSENSOR WITH MICRODIALYSIS
    PFEIFFER, EF
    MEYERHOFF, C
    BISCHOF, F
    KECK, FS
    KERNER, W
    [J]. HORMONE AND METABOLIC RESEARCH, 1993, 25 (02) : 121 - 124
  • [7] PFEIFFER EF, 1991, DIABETES 1991, P919
  • [8] GLUCOSE-CONCENTRATION IN SUBCUTANEOUS EXTRACELLULAR-SPACE
    SCHMIDT, FJ
    SLUITER, WJ
    SCHOONEN, AJM
    [J]. DIABETES CARE, 1993, 16 (05) : 695 - 700
  • [9] WERNER W, 1970, FRESEN Z ANAL CHEM, V252, P224