EFFECT OF SERUM SEPARATOR TUBES ON FREE AND TOTAL PHENYTOIN AND CARBAMAZEPINE SERUM CONCENTRATIONS

被引:11
作者
MAURO, LS [1 ]
MAURO, VF [1 ]
机构
[1] MED COLL HOSP,DEPT PHARM SERV,TOLEDO,OH
关键词
PHENYTOIN; CARBAMAZEPINE; SERUM SEPARATOR TUBE; BLOOD COLLECTION;
D O I
10.1097/00007691-199105000-00009
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The effect of serum separator tubes (SSTs) on free and total serum phenytoin and carbamazepine concentrations was determined by comparing standard no-additive tubes with SSTs (Becton Dickinson SST and Terumo Autosep). The influence of time prior to centrifugation, sample volume, and initial drug concentration on the effects were also studied. Results were analyzed using repeated measures two-way analysis of variance with tube type and either time, sample volume, or concentration as main effects. The most significant reductions noted were with Becton Dickinson SSTs in free and total serum phenytoin and total carbamazepine concentrations, where all reductions were < 10%. The only factor to significantly influence extent of reduction was the effect of time on total serum phenytoin concentration in Becton Dickinson SSTs. Terumo Autosep tubes caused no major reductions in free or total phenytoin or carbamazepine serum concentrations. Autosep tubes should provide accurate measurements of total and free serum phenytoin and carbamazepine concentrations. With Becton Dickinson SSTs, the reductions noted in free and total phenytoin and total carbamazepine concentrations were not large enough to preclude their clinical use. Becton Dickinson SSTs should not be used for determining free or total phenytoin or total carbamazepine concentrations for purposes of research.
引用
收藏
页码:240 / 243
页数:4
相关论文
共 6 条
[1]  
BERGQVIST Y, 1984, CLIN CHEM, V30, P465
[2]  
FITZSIMMONS WE, 1986, CLIN PHARMACY, V5, P923
[3]   SUITABILITY OF COLLECTION TUBES WITH SEPARATOR GELS FOR THERAPEUTIC DRUG-MONITORING [J].
KOCH, TR ;
PLATOFF, G .
THERAPEUTIC DRUG MONITORING, 1990, 12 (03) :277-280
[4]   BLOOD COLLECTION TUBES FOR TRICYCLIC ANTIDEPRESSANT DRUGS - A REEVALUATION [J].
ORSULAK, PJ ;
SINK, M ;
WEED, J .
THERAPEUTIC DRUG MONITORING, 1984, 6 (04) :444-448
[5]   STABILITY OF PHENYTOIN IN BLOOD COLLECTED IN VACUUM BLOOD COLLECTION TUBES [J].
PARISH, RC ;
ALEXANDER, T .
THERAPEUTIC DRUG MONITORING, 1990, 12 (01) :85-89
[6]   EFFECT OF SERUM SEPARATOR BLOOD COLLECTION TUBES ON DRUG CONCENTRATIONS [J].
QUATTROCCHI, F ;
KARNES, HT ;
ROBINSON, JD ;
HENDELES, L .
THERAPEUTIC DRUG MONITORING, 1983, 5 (03) :359-362