PULSED ELECTROCHEMICAL DETECTION OF PENICILLINS USING 3-STEP AND 4-STEP WAVE-FORMS

被引:21
作者
ALTUNATA, S [1 ]
EARLEY, RL [1 ]
MOSSMAN, DM [1 ]
WELCH, LE [1 ]
机构
[1] KNOX COLL,DEPT CHEM,GALESBURG,IL 61401
关键词
D O I
10.1016/0039-9140(94)00205-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Aided by the construction of a custom potentiostat, a series of different PAD waveforms were compared to find the optimum detector for penicillin oxidation. The waveforms included standard 3-step direct and indirect PAD in addition to reverse-PAD and 4-step PAD. Two new waveforms, the indirect reverse-PAD and the 4-step indirect PAD were examined in the study. Under the solvent conditions of the study (0.01M acetate buffer, pH 4.6) the indirect waveforms yielded the best detectability for penicillin G while the reverse-PAD waveforms yielded the worst performance. The 4-step PAD methods did not improve detectability when compared to the 3-step types, but they did provide output peak profiles with better shapes and less tailing. Although indirect waveforms gave better detectability than direct detection in the 0.01M acetate buffer solution, the limits of detection for each were found to be differing functions of ionic strength. At higher acetate concentrations, direct PAD was more favorable than indirect detection.
引用
收藏
页码:17 / 25
页数:9
相关论文
共 24 条
[1]  
Adams R. N., 1969, ELECTROCHEMISTRY SOL
[2]   ELECTROCATALYSIS AT PLATINUM-ELECTRODES FOR ANODIC ELECTROANALYSIS [J].
AUSTIN, DS ;
POLTA, JA ;
POLTA, TZ ;
TANG, APC ;
CABELKA, TD ;
JOHNSON, DC .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1984, 168 (1-2) :227-248
[3]   PULSED AMPEROMETRIC DETECTION BASED ON DIRECT AND INDIRECT ANODIC REACTIONS - A REVIEW [J].
AUSTINHARRISON, DS ;
JOHNSON, DC .
ELECTROANALYSIS, 1989, 1 (03) :189-197
[4]   AMPEROMETRIC DETECTION OF SIMPLE CARBOHYDRATES AT PLATINUM-ELECTRODES IN ALKALINE-SOLUTIONS BY APPLICATION OF A TRIPLE-PULSE POTENTIAL WAVEFORM [J].
HUGHES, S ;
JOHNSON, DC .
ANALYTICA CHIMICA ACTA, 1981, 132 (DEC) :11-22
[5]   AMPEROMETRIC DETECTION OF SIMPLE ALCOHOLS IN AQUEOUS-SOLUTIONS BY APPLICATION OF A TRIPLE-PULSE POTENTIAL WAVEFORM AT PLATINUM-ELECTRODES [J].
HUGHES, S ;
MESCHI, PL ;
JOHNSON, DC .
ANALYTICA CHIMICA ACTA, 1981, 132 (DEC) :1-10
[6]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION WITH TRIPLE-PULSE AMPEROMETRIC DETECTION OF CARBOHYDRATES IN BEVERAGES [J].
HUGHES, S ;
JOHNSON, DC .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1982, 30 (04) :712-714
[7]   LIQUID-CHROMATOGRAPHY WITH PULSED ELECTROCHEMICAL DETECTION AT GOLD AND PLATINUM-ELECTRODES [J].
JOHNSON, DC ;
LACOURSE, WR .
ANALYTICAL CHEMISTRY, 1990, 62 (10) :A589-A597
[8]  
JOHNSON DC, 1986, ANAL CHIM ACTA, V132, P187
[9]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION AND ELECTROCHEMICAL DETECTION OF PENICILLINS [J].
KIRCHMANN, E ;
WELCH, LE .
JOURNAL OF CHROMATOGRAPHY, 1993, 633 (1-2) :111-118
[10]   THE ELECTROCHEMICAL DETECTION OF PENICILLINS IN MILK [J].
KIRCHMANN, E ;
EARLEY, RL ;
WELCH, LE .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1994, 17 (08) :1755-1772