PRECISION AND ACCURACY FOR RAT WHOLE-BODY AND FEMUR BONE-MINERAL DETERMINATION WITH DUAL X-RAY ABSORPTIOMETRY

被引:35
作者
HAGIWARA, S
LANE, N
ENGELKE, K
SEBASTIAN, A
KIMMEL, DB
GENANT, HK
机构
[1] UNIV CALIF SAN FRANCISCO,DEPT RADIOL,505 PARNASSUS AVE,SAN FRANCISCO,CA 94143
[2] UNIV CALIF SAN FRANCISCO,GEN CLIN RES CTR,SAN FRANCISCO,CA 94143
[3] CREIGHTON UNIV,DEPT MED,OMAHA,NE 68151
来源
BONE AND MINERAL | 1993年 / 22卷 / 01期
关键词
DUAL X-RAY ABSORPTIOMETRY (DXA); RAT FEMUR; ASH WEIGHT; PRECISION; ACCURACY; BONE MINERAL CONTENT (BMC); BONE MINERAL DENSITY (BMD);
D O I
10.1016/S0169-6009(08)80081-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Precision and accuracy for rat whole body and excised femur bone mineral density (BMD) measurements were evaluated with two dual X-ray absorptiometry (DXA) systems. The 'small subject' mode on the Norland XR-26 (XR), and the 'ultra high resolution mode' on the Hologic QDR-1000/W(QDR) were used for the analysis. The whole body mode was only available on the XR. The lowest precision error for the whole body was found using a scan resolution of 1.0 x 1.0 mm and a scan speed of 15 mm/s. The scan spatial resolution of the femur measurement was approximately eight times higher on the QDR than on the XR. However, the XR allowed analysis of an arbitrary region of interest within the femur, which was not easily done with the QDR. Precision for the total femur measurement on the QDR (0.5-0.9%) was approximately two to four times superior to that of the XR (1.5-4.3%). The difference may be due to the superior scan resolution of the QDR. Bone mineral content and BMD on the QDR significantly declined with an increase of water depth (P < 0.001). No significant change was observed on the XR. Both DXA systems demonstrated an excellent correlation (r greater-than-or-equal-to 0.98) with ash weight under the scan conditions examined. The optimal scan condition for the excised femur measurement on the XR was obtained with 1.5-2.5 cm of perspex or water and a scan speed of 10 nun/s. For the QDR, we recommend scanning the rat femur with approximately 2.5 cm of perspex or water in terms of precision and accuracy.
引用
收藏
页码:57 / 68
页数:12
相关论文
共 17 条
[1]  
AMMANN P, 1992, J BONE MINER RES, V7, P311
[2]  
ANCILL A, 1992, J BONE MINERR RES, V7, pS278
[3]   EFFECT OF LONG-TERM ESTROGEN ADMINISTRATION ON BONE METABOLISM IN THE FEMALE RAT [J].
CRUESS, RL ;
HONG, KC .
ENDOCRINOLOGY, 1979, 104 (04) :1188-1193
[4]  
GENANT HK, 1990, OSTEOPOROSIS 1990, P435
[5]   THE EFFECT OF LACTATION ON THE MINERAL DISTRIBUTION PROFILE OF THE RAT FEMUR BY SINGLE PHOTON-ABSORPTIOMETRY [J].
HAGAMAN, JR ;
SANCHEZ, TV ;
MYERS, RC .
BONE, 1985, 6 (05) :301-305
[6]  
HANSEN MA, 1990, J NUCL MED, V31, P1156
[7]   QUANTITATIVE DIGITAL RADIOGRAPHY VERSUS DUAL PHOTON-ABSORPTIOMETRY OF THE LUMBAR SPINE [J].
KELLY, TL ;
SLOVIK, DM ;
SCHOENFELD, DA ;
NEER, RM .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1988, 67 (04) :839-844
[8]  
KIEBZAK GM, 1988, J BONE MINER RES, V3, P311
[9]   NONDESTRUCTIVE MEASUREMENT OF BONE-MINERAL IN FEMURS FROM OVARIECTOMIZED RATS [J].
KIMMEL, DB ;
WRONSKI, TJ .
CALCIFIED TISSUE INTERNATIONAL, 1990, 46 (02) :101-110