INDUCTION OF MICROCURRENTS IN CRITICALLY ILL PATIENTS IN MAGNETIC-RESONANCE SYSTEMS

被引:21
作者
PEDEN, CJ
COLLINS, AG
BUTSON, PC
WHITWAM, JG
YOUNG, IR
机构
[1] HAMMERSMITH HOSP,ROYAL POSTGRAD MED SCH,DEPT ANAESTHESIA,NMR UNIT,LONDON W12 0HS,ENGLAND
[2] GEC HIRST RES LABS,WEMBLEY,ENGLAND
关键词
MAGNETIC RESONANCE; VENTRICULAR FIBRILLATION; CARDIAC PACING; CATHETERIZATION; CENTRAL; VENOUS; PULMONARY ARTERY; CRITICAL CARE; ELECTROCARDIOGRAPHY;
D O I
10.1097/00003246-199312000-00021
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To determine whether electric current can be induced in intracardiac catheters, thermistor wires and pacing electrodes in patients during magnetic resonance imaging (MRI). Design: Prospective laboratory study. Setting: Postgraduate medical school hospital. Subjects: A sheep heart model. Interventions: None. Measurements and Main Results: Voltage generated by saline 0.9% flowing through a magnetic field and distribution of current from a catheter tip within a sheep heart model were measured in a 0.15 Tesla MRI system. Resistance of loops formed by pacing wires, a pacing electrode, and a thermistor wire were measured in saline 0.9%. Effects of rapidly changing magnetic fields and the movement of the beating heart on epicardial pacing wires were calculated theoretically. A flow of 200 mL/min of saline 0.9% induced a current of 0.1 microampere (mu A) (at 0.15 Tesla). From magnetic resonance images we derived a current density of similar to 0.004 mu A/mm(2) (at 0.15 Tesla). Internal resistance of pacing catheters and thermistor wires was >1 megaohm (M Omega). The maximum currents calculated (for a higher field strength of 1.5 Tesla) in a circuit formed by epicardial pacing wires were 80 mu A, induced by the beating heart moving the wires through the magnetic field and 46 mu A, induced by the rapidly changing magnetic fields. Conclusions: Current generated by flow of conducting fluid should be safe. Pacing catheters and thermistor wires should be safe if well insulated and disconnected from external electric connections. However, current induced in epicardial pacing wires may be a hazard, and precautions should be taken. External wire tips must be separated, insulated, and coiled to lie along the axis of the magnetic field. Electrocardiography is required, and defibrillation equipment should be available.
引用
收藏
页码:1923 / 1928
页数:6
相关论文
共 15 条
[1]   THRESHOLDS FOR PHYSIOLOGICAL-EFFECTS DUE TO RF AND MAGNETIC-FIELDS USED IN NMR IMAGING [J].
BUDINGER, TF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (02) :2821-2825
[2]   THE EFFECT OF MAGNETIC-SUSCEPTIBILITY VARIATIONS IN NMR IMAGING AND NMR-SPECTROSCOPY INVIVO [J].
COX, IJ ;
BYDDER, GM ;
GADIAN, DG ;
YOUNG, IR ;
PROCTOR, E ;
WILLIAMS, SR ;
HART, I .
JOURNAL OF MAGNETIC RESONANCE, 1986, 70 (01) :163-168
[3]   CHANGES IN CARDIAC EXCITABILITY AND VULNERABILITY IN NMR FIELDS [J].
DOHERTY, JU ;
WHITMAN, GJR ;
ROBINSON, MD ;
HARKEN, AH ;
SIMSON, MB ;
SPEAR, JF ;
JOSEPHSON, ME .
INVESTIGATIVE RADIOLOGY, 1985, 20 (02) :129-135
[4]  
GREEN HL, 1972, BIO MED ENG, V10, P408
[5]   NUCLEAR MAGNETIC-RESONANCE MEASUREMENTS - CLINICAL-APPLICATIONS [J].
GUTIERREZ, G ;
ANDRY, JM .
CRITICAL CARE MEDICINE, 1989, 17 (01) :73-82
[6]  
JOY M, 1988, MAGN RESON IMAGING, V7, P89
[7]   SAFETY CONSIDERATIONS IN MR IMAGING [J].
KANAL, E ;
SHELLOCK, FG ;
TALAGALA, L .
RADIOLOGY, 1990, 176 (03) :593-606
[8]   BURNS ASSOCIATED WITH CLINICAL MR EXAMINATIONS [J].
KANAL, E ;
SHELLOCK, FG .
RADIOLOGY, 1990, 175 (02) :585-585
[9]   THRESHOLDS OF FIBRILLATING LEAKAGE CURRENTS ALONG INTRACARDIAC CATHETERS - EXPERIMENTAL STUDY [J].
LEE, WR ;
SCOTT, JR .
CARDIOVASCULAR RESEARCH, 1973, 7 (04) :495-500
[10]   MAGNETIC-RESONANCE FOR THE ANESTHETIST .1. PHYSICAL PRINCIPLES, APPLICATIONS, SAFETY ASPECTS [J].
MENON, DK ;
PEDEN, CJ ;
HALL, AS ;
SARGENTONI, J ;
WHITWAM, JG .
ANAESTHESIA, 1992, 47 (03) :240-255