COMPARISON OF THE EFFECT OF DIFFERENT REFERENCE DATA ON LUNAR-DPX AND HOLOGIC-QDR-1000 DUAL-ENERGY X-RAY ABSORPTIOMETERS

被引:49
作者
LASKEY, MA
CRISP, AJ
COLE, TJ
COMPSTON, JE
机构
[1] ADDENBROOKES HOSP,DEPT NUCL MED,BONE DENS UNIT,CAMBRIDGE CB2 2QQ,ENGLAND
[2] ADDENBROOKES HOSP,DEPT RHEUMATOL,CAMBRIDGE CB2 2QQ,ENGLAND
[3] UNIV CAMBRIDGE,ADDENBROOKES HOSP,SCH CLIN,DEPT MED,CAMBRIDGE CB2 2QQ,ENGLAND
关键词
DUAL-ENERGY X-RAY ABSORPTIOMETRY; REFERENCE DATA; LUNAR-DPX; HOLOGIC-QDR-1000;
D O I
10.1259/0007-1285-65-780-1124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have investigated whether the Lunar DPX (software 3.4) and Hologic QDR-1000 dual-energy X-ray absorptiometers have comparable normal reference databases for the spine and femur of white UK and USA subjects. After conversion for systematic differences in absolute bone density values between the two systems, the reference databases were very similar for the spine in young subjects, but there were clear differences in the femur databases of young females and males of all ages. These differences were confirmed by comparing the percent age-matched and young values determined by the two systems for subjects scanned on both systems. Thus the diagnosis and management of a patient could differ, depending on the system used for the bone density measurements.
引用
收藏
页码:1124 / 1129
页数:6
相关论文
共 15 条
[1]   THE INTERCHANGEABILITY OF RADIOISOTOPE AND X-RAY BASED MEASUREMENTS OF BONE-MINERAL DENSITY [J].
ADACHI, JD ;
WEBBER, CE .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (759) :217-220
[2]   INTERSITE COMPARISON OF THE HOLOGIC QDR-1000 DUAL ENERGY X-RAY BONE DENSITOMETER [J].
BLAKE, GM ;
TONG, CM ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (761) :440-446
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   X-RAY DUAL-PHOTON ABSORPTIOMETRY - A NEW METHOD FOR THE MEASUREMENT OF BONE-DENSITY [J].
CULLUM, ID ;
ELL, PJ ;
RYDER, JP .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (739) :587-592
[5]   DUAL-ENERGY RADIOGRAPHIC ABSORPTIOMETRY OF THE LUMBAR SPINE - CLINICAL-EXPERIENCE WITH 2 DIFFERENT SYSTEMS [J].
GUNDRY, CR ;
MILLER, CW ;
RAMOS, E ;
MOSCONA, A ;
STEIN, JA ;
MAZESS, RB ;
SARTORIS, DJ ;
RESNICK, D .
RADIOLOGY, 1990, 174 (02) :539-541
[6]  
KELLY TL, 1990, J BONE MINER RES S2, V5, P702
[7]   COMPARATIVE PERFORMANCE INVITRO AND INVIVO OF LUNAR DPX AND HOLOGIC QDR-1000 DUAL ENERGY X-RAY ABSORPTIOMETERS [J].
LASKEY, MA ;
FLAXMAN, ME ;
BARBER, RW ;
TRAFFORD, S ;
HAYBALL, MP ;
LYTTLE, KD ;
CRISP, AJ ;
COMPSTON, JE .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (767) :1023-1029
[8]   PERFORMANCE EVALUATION OF A DUAL-ENERGY X-RAY BONE DENSITOMETER [J].
MAZESS, R ;
COLLICK, B ;
TREMPE, J ;
BARDEN, H ;
HANSON, J .
CALCIFIED TISSUE INTERNATIONAL, 1989, 44 (03) :228-232
[9]  
MAZESS RB, 1991, J BONE MINER RES, V6, P799
[10]  