INVESTIGATIONS ON THE APPLICABILITY OF SUPERHEATED DROP DETECTORS IN NEUTRON DOSIMETRY

被引:43
作者
APFEL, RE
ROY, SC
机构
关键词
DOSIMETRY;
D O I
10.1016/0167-5087(84)90234-5
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Superheated drop detectors (SDD) made of four different liquids have been tested at the US National Bureau of Standards to study their response under irradiation by neutrons ranging in energy from fast to thermal. A comparison of responses and features of SDDs with the ideal neutron dosimeter and other currently used dosimeters has been made. The results, among other observations, show that the SDD made of refrigerant 12 (CCl//2F//2), du Pont Trade Name Freon 12, is a promising candidate for use in neutron dosimetry.
引用
收藏
页码:582 / 587
页数:6
相关论文
共 15 条
[11]  
GRIFFITH RV, 1981, 8TH DOE WORKSH PERS
[12]   CALCULATED SPECTRA FOR DOSIMETRY OF D2O-MODERATED NEUTRONS [J].
ING, H ;
CROSS, WG .
HEALTH PHYSICS, 1977, 32 (05) :351-357
[13]  
OSWALD RA, 1977, ANN M HLTH PHYSICS S
[14]  
Peyrou Ch., 1967, BUBBLE SPARK CHAMBER, V1
[15]   ON THE THEORY OF THE BUBBLE CHAMBER [J].
SEITZ, F .
PHYSICS OF FLUIDS, 1958, 1 (01) :2-13