HIGHLY SENSITIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOR THE MEASUREMENT OF MALONDIALDEHYDE IN BIOLOGICAL SAMPLES

被引:85
作者
FUKUNAGA, K [1 ]
SUZUKI, T [1 ]
TAKAMA, K [1 ]
机构
[1] HOKKAIDO UNIV, FAC FISHERIES, DEPT FOOD SCI & TECHNOL, HAKODATE, HOKKAIDO 041, JAPAN
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1993年 / 621卷 / 01期
关键词
D O I
10.1016/0378-4347(93)80078-I
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive method for the measurement of malondialdehyde as thiobarbituric acid-malondialdehyde complex by reversed-phase high-performance liquid chromatography with fluorescence detection in biological samples is described. As samples, 20 mu l of rat plasma or 10% (w/v) liver homogenate mixed with 2.0% thiobarbituric acid in 2 M sodium acetate buffer containing 0.05% butyl hydroxytoluene (pH 3.5) were heated at 95 degrees C for 45 min to give the complex. The complex, extracted with n-butanol, was chromatographed on a system equipped with a reversed-phase column, and the eluted peak was monitored with a fluorescence detector (excitation wavelength 515 nm, emission wavelength 553 nm). The mobile phase was a acetonitrile-water (2:8, v/v) under isocratic conditions at ambient temperature, and a single analysis was done in ca. 4 min. The minimum detection level for malondialdehyde was as low as 0.05 pmol. The n-butanol extract was stable at least for 3 days. The simple mobile phase, the extremely sensitive detection limit, and the stability of the complex make this system applicable to routine clinical analysis with a small amount of tissue or biopsy sample.
引用
收藏
页码:77 / 81
页数:5
相关论文
共 22 条