THE EFFECT OF RESPIRATION ON THE CONTRAST AND SHARPNESS OF LIVER-LESIONS IN MRI

被引:23
作者
BUTTS, K [1 ]
RIEDERER, SJ [1 ]
EHMAN, RL [1 ]
机构
[1] MAYO CLIN & MAYO FDN,MAGNET RESONANCE LAB,ROCHESTER,MN 55905
关键词
FAST SCANNING; MR IMAGING; LIVER IMAGING;
D O I
10.1002/mrm.1910330102
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work demonstrates the effects of through-plane motion due to respiration on contrast and sharpness of liver lesions in MRI. The effects of slice coverage with and without such respiratory motion is also reported. This work is comprised of two parts: a theoretical prediction of liver-lesion contrast and blur with and without respiration and an experimental validation using gel phantoms of the predicted results. Both theory and experiment show a loss of contrast, increasing with amplitude of the peak-to-peak motion. The loss of contrast for a 5-mm lesion at normal respiration of 15 mm peak-to-peak superior-inferior motion is approximate to 10% with a low order sorted respiratory ordered phase encoding acquisition and approximate to 50% for an unsorted acquisition. Lesion blur is greatest for the low order sorted acquisition while the unsorted and high sort acquisitions maintain edge definition, Breath-hold imaging is potentially superior to nonbreath-hold imaging in liver lesion contrast and edge definition, but is more sensitive to inadequate slice coverage.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 20 条
  • [1] RESPIRATORY ORDERED PHASE ENCODING (ROPE) - A METHOD FOR REDUCING RESPIRATORY MOTION ARTIFACTS IN MR IMAGING
    BAILES, DR
    GILDERDALE, DJ
    BYDDER, GM
    COLLINS, AG
    FIRMIN, DN
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) : 835 - 838
  • [2] INTERLEAVED ECHO-PLANAR IMAGING ON A STANDARD MRI SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    THOMPSON, RM
    JACK, CR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) : 67 - 72
  • [3] ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    FELMLEE, JP
    GRIMM, RC
    [J]. RADIOLOGY, 1993, 189 (01) : 259 - 264
  • [4] SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST
    EDELMAN, RR
    WALLNER, B
    SINGER, A
    ATKINSON, DJ
    SAINI, S
    [J]. RADIOLOGY, 1990, 177 (02) : 515 - 521
  • [5] EHMAN EL, 1987, RADIOLOGY, V159, P777
  • [6] SPATIAL PRESATURATION - A METHOD FOR SUPPRESSING FLOW ARTIFACTS AND IMPROVING DEPICTION OF VASCULAR ANATOMY IN MR IMAGING
    FELMLEE, JP
    EHMAN, RL
    [J]. RADIOLOGY, 1987, 164 (02) : 559 - 564
  • [7] PSEUDO-GATING - ELIMINATION OF PERIODIC MOTION ARTIFACTS IN MAGNETIC-RESONANCE-IMAGING WITHOUT GATING
    HAACKE, EM
    LENZ, GW
    NELSON, AD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (02) : 162 - 174
  • [8] HAASE A, 1985, PHYS MED BIOL, V36, P341
  • [9] HOLSINGERBAMPTO.AE, 1991, RADIOLOGY, V181, P25
  • [10] RESPIRATORY KINEMATICS OF THE UPPER ABDOMINAL ORGANS - A QUANTITATIVE STUDY
    KORIN, HW
    EHMAN, RL
    RIEDERER, SJ
    FELMLEE, JP
    GRIMM, RC
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) : 172 - 178