HARMONIC ENHANCEMENT OF A DETECTED ION-CYCLOTRON RESONANCE SIGNAL BY USE OF SEGMENTED DETECTION ELECTRODES

被引:18
作者
LIMBACH, PA [1 ]
GROSSHANS, PB [1 ]
MARSHALL, AG [1 ]
机构
[1] OHIO STATE UNIV,DEPT CHEM,120 W 18TH AVE,COLUMBUS,OH 43210
来源
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES | 1993年 / 123卷 / 01期
关键词
FT-ICR; TRAP DESIGN; RELAXATION MECHANISMS; ICR RADIUS;
D O I
10.1016/0168-1176(93)87052-T
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
For certain relaxation mechanisms, FT-ICR mass resolving power increases with harmonic order (e.g., threefold enhancement in mass resolving power for the third harmonic frequency (3omega+) compared to fundamental frequency (omega+)). Moreover, ion cyclotron orbital radius may be determined from the ratio of the magnitude of the 3omega+ signal to that of omega+, M(3omega+)/M(omega+). Unfortunately, the magnitude of the 3omega+ signal from a standard cubic FT-ICR ion trap is small (usually less-than-or-equal-to 10% of the fundamental). In this paper, we present a new theoretical trap design for increasing the 3omega+ content of the detected signal, and we demonstrate experimentally (by measurement of the ICR orbital radius) a wiring configuration which substantially enhances (factor of up to almost-equal-to 5) the 3omega+ signal.