PORTAL DOSE IMAGES .2. PATIENT DOSE ESTIMATION

被引:59
作者
YING, XR
GEER, LY
WONG, JW
机构
[1] WASHINGTON UNIV,MALLINCKRODT INST RADIOL,PHYS SECT,510 S KINGSHIGHWAY BLVD,ST LOUIS,MO 63110
[2] CHINESE ACAD SCI,INST AUTOMAT,BEIJING,PEOPLES R CHINA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1990年 / 18卷 / 06期
关键词
On-line radiotherapy imaging; Patient dose estimation; Treatment uncertainties; Treatment verification;
D O I
10.1016/0360-3016(90)90323-C
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Due to the many sources of uncertainties in radiotherapy, conventional treatment planning can only provide a nominal presentation of the dose delivered to the patient. Provided that large setup errors can be detected and corrected, we propose that measured and calculated portal dose images can be used to improve estimation of patient dose. The iterative approach described in this paper requires an accurate method of dose calculations, 3-dimensional CT data that closely represents the patient and the measured portal dose image. From the CT data, a portal dose image is calculated for comparison with the measured one. The differences are then used to modify the original CT data so that a new image can be calculated. The process is repeated until the calculated and measured images agree to satisfaction. At that point, the internal dose is calculated using the modified CT data. If the method is successful, daily portal dose images could be used to cumulatively estimate patient dose throughout the course of treatment. This manuscript describes 60Co simulation results to demonstrate the feasibility of the approach. © 1990.
引用
收藏
页码:1465 / 1475
页数:11
相关论文
共 22 条
[1]  
Fraass B. A., 1987, USE COMPUTERS RAD TH, P273
[2]   MULTIDIMENSIONAL TREATMENT PLANNING .2. BEAM EYE-VIEW, BACK PROJECTION, AND PROJECTION THROUGH CT SECTIONS [J].
GOITEIN, M ;
ABRAMS, M ;
ROWELL, D ;
POLLARI, H ;
WILES, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (06) :789-797
[3]   MULTIDIMENSIONAL TREATMENT PLANNING .1. DELINEATION OF ANATOMY [J].
GOITEIN, M ;
ABRAMS, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (06) :777-787
[4]  
GOITEIN M, 1985, MED PHYS, V12, P601
[5]   ACCURACY IN RADIATION-FIELD ALIGNMENT IN HEAD AND NECK-CANCER - A PROSPECTIVE-STUDY [J].
HUIZENGA, H ;
LEVENDAG, PC ;
DEPORRE, PMZR ;
VISSER, AG .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (02) :181-187
[6]  
KINZIE JJ, 1983, CANCER, V52, P2223, DOI 10.1002/1097-0142(19831215)52:12<2223::AID-CNCR2820521209>3.0.CO
[7]  
2-D
[8]   AN ONLINE ELECTRONIC PORTAL IMAGING-SYSTEM FOR EXTERNAL BEAM RADIOTHERAPY [J].
LAM, KS ;
PARTOWMAH, M ;
LAM, WC .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (706) :1007-1013
[9]  
LASCHINGER JC, 1987, J MED IMAGING, V1, P228
[10]   USE OF DIGITAL FLUOROSCOPY AS AN ONLINE VERIFICATION DEVICE IN RADIATION-THERAPY [J].
LEONG, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (09) :985-992