PORTAL DOSE IMAGES .2. PATIENT DOSE ESTIMATION

被引:59
作者
YING, XR
GEER, LY
WONG, JW
机构
[1] WASHINGTON UNIV,MALLINCKRODT INST RADIOL,PHYS SECT,510 S KINGSHIGHWAY BLVD,ST LOUIS,MO 63110
[2] CHINESE ACAD SCI,INST AUTOMAT,BEIJING,PEOPLES R CHINA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1990年 / 18卷 / 06期
关键词
On-line radiotherapy imaging; Patient dose estimation; Treatment uncertainties; Treatment verification;
D O I
10.1016/0360-3016(90)90323-C
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Due to the many sources of uncertainties in radiotherapy, conventional treatment planning can only provide a nominal presentation of the dose delivered to the patient. Provided that large setup errors can be detected and corrected, we propose that measured and calculated portal dose images can be used to improve estimation of patient dose. The iterative approach described in this paper requires an accurate method of dose calculations, 3-dimensional CT data that closely represents the patient and the measured portal dose image. From the CT data, a portal dose image is calculated for comparison with the measured one. The differences are then used to modify the original CT data so that a new image can be calculated. The process is repeated until the calculated and measured images agree to satisfaction. At that point, the internal dose is calculated using the modified CT data. If the method is successful, daily portal dose images could be used to cumulatively estimate patient dose throughout the course of treatment. This manuscript describes 60Co simulation results to demonstrate the feasibility of the approach. © 1990.
引用
收藏
页码:1465 / 1475
页数:11
相关论文
共 22 条
[11]  
LEONG J, 1987, PHYS MED BIOL, V32, P307
[12]   A CONVOLUTION METHOD OF CALCULATING DOSE FOR 15-MV X-RAYS [J].
MACKIE, TR ;
SCRIMGER, JW ;
BATTISTA, JJ .
MEDICAL PHYSICS, 1985, 12 (02) :188-196
[13]  
MARKS JE, 1976, CANCER, V37, P2755, DOI 10.1002/1097-0142(197606)37:6<2755::AID-CNCR2820370628>3.0.CO
[14]  
2-J
[15]   DIFFERENTIAL PENCIL BEAM DOSE COMPUTATION MODEL FOR PHOTONS [J].
MOHAN, R ;
CHUI, C ;
LIDOFSKY, L .
MEDICAL PHYSICS, 1986, 13 (01) :64-73
[16]  
PURDY JA, 1987, 9TH INT C US COMP RA, P277
[17]   ACCURACY OF RADIATION-FIELD ALIGNMENT IN CLINICAL-PRACTICE [J].
RABINOWITZ, I ;
BROOMBERG, J ;
GOITEIN, M ;
MCCARTHY, K ;
LEONG, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (10) :1857-1867
[18]   A MATRIX IONIZATION-CHAMBER IMAGING DEVICE FOR ONLINE PATIENT SETUP VERIFICATION DURING RADIOTHERAPY [J].
VANHERK, M ;
MEERTENS, H .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (04) :369-378
[19]   PORTAL DOSE IMAGES .1. QUANTITATIVE TREATMENT PLAN VERIFICATION [J].
WONG, JW ;
SLESSINGER, ED ;
HERMES, RE ;
OFFUTT, CJ ;
ROY, T ;
VANNIER, MW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (06) :1455-1463
[20]  
WONG JW, 1988, RADIOTHER SYSTEM RES, V5, P213