FREQUENCY-SPECTRA OF VENTRICULAR-TACHYCARDIA AND SINUS RHYTHM IN HUMAN INTRACARDIAC ELECTROGRAMS - APPLICATION TO TACHYCARDIA DETECTION FOR CARDIAC-PACEMAKERS
被引:28
作者:
PANNIZZO, F
论文数: 0引用数: 0
h-index: 0
机构:MONTEFIORE MED CTR,DEPT SURG,BRONX,NY 10467
PANNIZZO, F
FURMAN, S
论文数: 0引用数: 0
h-index: 0
机构:MONTEFIORE MED CTR,DEPT SURG,BRONX,NY 10467
FURMAN, S
机构:
[1] MONTEFIORE MED CTR,DEPT SURG,BRONX,NY 10467
[2] YESHIVA UNIV ALBERT EINSTEIN COLL MED,BRONX,NY 10461
MATHEMATICAL TRANSFORMATIONS - Fast Fourier Transforms - SIGNAL FILTERING AND PREDICTION - Medical Applications - SPECTRUM ANALYSIS;
D O I:
10.1109/10.2111
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
Presently available antitachycardia pacemakers detect ventricular tachycardia (VT) by measurement of the period of the intracardiac electrogram (EGM). The resulting inability to differentiate between VT and rapid normal sinus rhythm (NSR) can cause inappropriate pacemaker output, possibly initiating an arrhythmia where none existed previously. It is reported that FFT analysis of matched-pair NSR-VT EGM recordings from 33 patients revealed a mean ( plus or minus S. D. ) NSR-VT difference in the peak amplitude point of 8 ( plus or minus 7) Hz and a difference of 18 ( plus or minus 18) Hz in the minus 3 dB point. NSR-VT amplitude differences could be significantly increased by filtering over the range of greatest spectral differences. Variable-passband programmable filters could enhance VT detection in antitachycardia pacemakers.