A MODERATE-POWER NITROGEN MICROWAVE-INDUCED PLASMA AS AN ALTERNATIVE ION-SOURCE FOR MASS-SPECTROMETRY

被引:40
作者
SHEN, WL [1 ]
DAVIDSON, TM [1 ]
CREED, JT [1 ]
CARUSO, JA [1 ]
机构
[1] UNIV CINCINNATI,DEPT CHEM,CINCINNATI,OH 45221
关键词
D O I
10.1366/0003702904086623
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The moderate-power nitrogen plasma is evaluated as an alternative ion source for plasma mass spectrometry. The nitrogen MIP-MS exhibits linear dynamic ranges of about 3-4 orders of magnitude and low ppb to sub-ppb detection limits for most elements under multielement scanning conditions. Nitrogen MIP-MS is capable of determining K, Ca, Cr, As, and Se at their major isotopes. In addition, nitrogen MIP-MS provides lower detection limits than argon ICP-MS for K (0.48 ppb) and Ca (0.24 ppb). The nitrogen MIP-MS system allows the determination of isotope ratios for 10 ppm of Ca (40/44) and 100 ppb of K (39/41), Cr (52/53), and Se (80/78) with less than 5% error.
引用
收藏
页码:1003 / 1010
页数:8
相关论文
共 20 条
[11]   INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
HOUK, RS ;
THOMPSON, JJ .
MASS SPECTROMETRY REVIEWS, 1988, 7 (04) :425-461
[12]   ALLEVIATION OF OVERLAP INTERFERENCES FOR DETERMINATION OF POTASSIUM ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
JIANG, SJ ;
HOUK, RS ;
STEVENS, MA .
ANALYTICAL CHEMISTRY, 1988, 60 (11) :1217-1221
[13]   DEVELOPMENT AND ASSESSMENT OF A HELIUM INDUCTIVELY COUPLED PLASMA IONIZATION SOURCE FOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
KOPPENAAL, DW ;
QUINTON, LF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :667-672
[14]   DETECTION OF HALOGENATED COMPOUNDS BY CAPILLARY GAS-CHROMATOGRAPHY WITH HELIUM PLASMA MASS-SPECTROMETRY DETECTION [J].
MOHAMAD, AH ;
CREED, JT ;
DAVIDSON, TM ;
CARUSO, JA .
APPLIED SPECTROSCOPY, 1989, 43 (07) :1127-1131
[15]   INDUCTIVELY COUPLED HELIUM PLASMA AS AN ION-SOURCE FOR MASS-SPECTROMETRY [J].
MONTASER, A ;
CHAN, SK ;
KOPPENAAL, DW .
ANALYTICAL CHEMISTRY, 1987, 59 (08) :1240-1242
[16]   DETECTION OF HALOGENS AS POSITIVE-IONS USING A HE MICROWAVE INDUCED PLASMA AS AN ION-SOURCE FOR MASS-SPECTROMETRY [J].
SATZGER, RD ;
FRICKE, FL ;
BROWN, PG ;
CARUSO, JA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (05) :705-712
[17]   BACKGROUND SPECTRAL FEATURES FOR MODERATE-POWER NITROGEN MICROWAVE-INDUCED PLASMA MASS-SPECTROMETRY [J].
SHEN, WL ;
DAVIDSON, TM ;
CREED, JT ;
CARUSO, JA .
APPLIED SPECTROSCOPY, 1990, 44 (06) :1011-1014
[18]   BACKGROUND SPECTRAL FEATURES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
TAN, SH ;
HORLICK, G .
APPLIED SPECTROSCOPY, 1986, 40 (04) :445-460
[19]   OXIDE, HYDROXIDE, AND DOUBLY CHARGED ANALYTE SPECIES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
VAUGHAN, MA ;
HORLICK, G .
APPLIED SPECTROSCOPY, 1986, 40 (04) :434-445
[20]   USE OF THE MICROWAVE-INDUCED NITROGEN DISCHARGE AT ATMOSPHERIC-PRESSURE AS AN ION-SOURCE FOR ELEMENTAL MASS-SPECTROMETRY [J].
WILSON, DA ;
VICKERS, GH ;
HIEFTJE, GM .
ANALYTICAL CHEMISTRY, 1987, 59 (13) :1664-1670