UNIVERSAL BONE CUTTING DEVICE FOR PRECISION KNEE REPLACEMENT ARTHROPLASTY AND OSTEOTOMY

被引:5
作者
COOKE, TDV
SAUNDERS, G
SIU, D
YOSHIOKA, Y
WEVERS, H
机构
[1] QUEENS UNIV,DEPT MECH ENGN,KINGSTON K7L 3N6,ONTARIO,CANADA
[2] KOBE UNIV,SCH MED,DEPT ORTHOPAED SURG,KOBE 650,JAPAN
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1985年 / 7卷 / 01期
基金
英国医学研究理事会;
关键词
BIOLOGICAL MATERIALS - Bone - BIOMEDICAL ENGINEERING - Surgery Applications - PROSTHETICS - Stability;
D O I
10.1016/0141-5425(85)90008-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Since malplacement contributes most to loosening of total knee replacements (TKR), a jig was devised, aligned to and mounted on th tibia, with a 3-degree of freedom sliding saw. A central distractor, attached to the jig, positions and aligns the knee at 0 degree or 90 degree . The femur is then rigidly linked to the jig for bone cutting. Resurfacing designs (Cloutier and Townley) have been regularly implanted, aligned plus or minus 1 degree . This precision should minimize loosening and improve function.
引用
收藏
页码:45 / 50
页数:6
相关论文
共 14 条
[1]
BARGREN JH, 1983, CLIN ORTHOPAEDICS, V173, P178
[2]
BARGREN JH, 1976, CLIN ORTHOP RELAT R, V120, P65
[3]
OSTEOGENIC PHENOMENA ACROSS ENDOSTEAL BONE-IMPLANT SPACES WITH POROUS SURFACED INTRA-MEDULLARY IMPLANTS [J].
BOBYN, JD ;
PILLIAR, RM ;
CAMERON, HU ;
WEATHERLY, GC .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :145-153
[4]
BRYANT JT, 1983, TECHNIQUE MEASURING
[5]
CAMERON HU, 1976, J BIOMED MATER RES, V10, P195
[7]
ESPIN E, 1983, SURVEY OSTEOTOMY SAW
[8]
HUNGERFORD DS, 1980, HOWMEDICA R UNIVERSA
[9]
INSALL J, 1979, CLIN ORTHOP RELAT R, V145, P68
[10]
MECHANICAL CAUSES OF LOOSENING IN KNEE-JOINT REPLACEMENT [J].
KAGAN, A .
JOURNAL OF BIOMECHANICS, 1977, 10 (07) :387-391