NONSTANDARD CT-SCANNERS - THEIR ROLE IN RADIOTHERAPY

被引:14
作者
WEBB, S [1 ]
机构
[1] ROYAL MARSDEN HOSP,LONDON,ENGLAND
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1990年 / 19卷 / 06期
基金
英国医学研究理事会;
关键词
CT-SCANNERS; IMAGING IN RADIOTHERAPY;
D O I
10.1016/0360-3016(90)90377-V
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
In the past 10 years a number of groups worldwide have investigated the extent to which the imaging requirements for radiotherapy planning may be met by the development of non-standard computed tomography (CT) scanners. Some of these groups have constructed apparatus based on a radiotherapy simulator gantry, others around a linac or a special purpose gantry. The performance of these systems is reviewed and the extent to which they have justified their promise has been assessed. The major use of simulator-based machines has been in planning radiotherapy in the thorax where tissue inhomogeneity corrections are important. Linac-based machines yield images useful for checking the positioning of patients at the time of therapy. Interest at the Royal Marsden Hospital culminated in the construction of a simulator-based CT system which achieved its present form in late 1985. The importance of the subsequent clinical study lies in our conclusive evidence that planning conservative radiotherapy for the breast without multiple-level CT performed in the treatment position can have implications for local control of disease, pulmonary damage, and cosmesis.
引用
收藏
页码:1589 / 1607
页数:19
相关论文
共 98 条