PSEUDOMONAS SEPTICEMIA DUE TO DEFICIENT DISINFECTANT MIXING DURING REUSE

被引:31
作者
VANHOLDER, R [1 ]
VANHAECKE, E [1 ]
RINGOIR, S [1 ]
机构
[1] LAB PHARMACEUT MICROBIOL & HYG,GHENT,BELGIUM
关键词
REUSE; SEPTICEMIA; BACTEREMIA; HEMODIALYSIS; PSEUDOMONAS;
D O I
10.1177/039139889201500104
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe a cluster of four septicemias with pseudomonas, that occurred in a unit performing formaldehyde reuse of capillary dialyzers. Samples of blood, heparin solutions, dialysate and effluent of reused dialyzers, were evaluated bacteriologically and upon the adequacy of the reuse procedure. Pseudomonas aeruginosa, vesicularis and/or xanthomonas maltophilia were found on the blood cultures obtained during the septicemic reactions, and in the effluent of two reprocessed dialyzers not yet used (> 10(4) CFU/ml). These two dialyzers had also extremely low formaldehyde concentrations (0.0014 and 0.005% versus the expected 4%). Membrane and antibiogram characteristics of a Pseudomonas aeruginosa strain, recovered from the blood cultures in one patient, and of a strain found in the effluent of one of the two contaminated reprocessed dialyzers, were the same. The problem was attributed to the inadequate mixing of the disinfectant with the tap water used in the automated reprocessing device, in the absence of an alarm disclosing this failure.
引用
收藏
页码:19 / 24
页数:6
相关论文
共 22 条
[1]   ACINETOBACTER-CALCOACETICUS OUTBREAK ASSOCIATED WITH PERITONEAL-DIALYSIS [J].
ABRUTYN, E ;
GOODHART, GL ;
ROOS, K ;
ANDERSON, R ;
BUXTON, A .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 1978, 107 (04) :328-335
[2]   REUSE OF HEMODIALYZERS - RESULTS OF NATIONWIDE SURVEILLANCE FOR ADVERSE-EFFECTS [J].
ALTER, MJ ;
FAVERO, MS ;
MILLER, JK ;
COLEMAN, PJ ;
BLAND, LA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 260 (14) :2073-2076
[3]  
BALTER P, 1979, INT J ARTIF ORGANS, V2, P112
[4]   BACTERIAL AND ENDOTOXIN PERMEABILITY OF HEMODIALYSIS MEMBRANES [J].
BERNICK, JJ ;
PORT, FK ;
FAVERO, MS ;
BROWN, DG .
KIDNEY INTERNATIONAL, 1979, 16 (04) :491-496
[5]  
BERNICK JJ, 1980, CLIN NEPHROL, V14, P13
[6]   INVIVO STUDIES OF DIALYSIS-RELATED ENDOTOXEMIA AND BACTEREMIA [J].
BERNICK, JJ ;
PORT, FK ;
FAVERO, MS .
NEPHRON, 1981, 27 (06) :307-312
[7]   INFECTIONS WITH MYCOBACTERIUM-CHELONEI IN PATIENTS RECEIVING DIALYSIS AND USING PROCESSED HEMODIALYZERS [J].
BOLAN, G ;
REINGOLD, AL ;
CARSON, LA ;
SILCOX, VA ;
WOODLEY, CL ;
HAYES, PS ;
HIGHTOWER, AW ;
MCFARLAND, L ;
BROWN, JW ;
PETERSEN, NJ ;
FAVERO, MS ;
GOOD, RC ;
BROOME, CV .
JOURNAL OF INFECTIOUS DISEASES, 1985, 152 (05) :1013-1019
[8]   WATER-QUALITY - A NEGLECTED PROBLEM IN HEMODIALYSIS [J].
BOMMER, J ;
RITZ, E .
NEPHRON, 1987, 46 (01) :1-6
[9]   HOST DEFENSES AND IMMUNOLOGICAL ALTERATIONS ASSOCIATED WITH CHRONIC-HEMODIALYSIS [J].
GOLDBLUM, SE ;
REED, WP .
ANNALS OF INTERNAL MEDICINE, 1980, 93 (04) :597-613
[10]   PYROGENIC REACTIONS ASSOCIATED WITH THE REUSE OF DISPOSABLE HOLLOW-FIBER HEMODIALYZERS [J].
GORDON, SM ;
TIPPLE, M ;
BLAND, LA ;
JARVIS, WR .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 260 (14) :2077-2081