TIBIAL HIGH-DENSITY POLYETHYLENE WEAR IN CONFORMING TIBIOFEMORAL PROSTHESES

被引:62
作者
PLANTEBORDENEUVE, P [1 ]
FREEMAN, MAR [1 ]
机构
[1] ROYAL LONDON HOSP,COLL MED,DEPT ORTHOPAED,BONE & JOINT RES UNIT,25-29 ASHFIELD ST,LONDON E1 1AD,ENGLAND
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1993年 / 75卷 / 04期
关键词
D O I
10.1302/0301-620X.75B4.8331121
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We have studied 27 tibial prostheses retrieved from knee replacements after 1 to 9 years. In 22 the femoral components were of cobalt-chrome, in five polyacetal. The design of the components gave a nominal contact area of 320 mm2 on each condyle. The tibial component was of high-density polyethylene (HDP) at least 6 mm thick, and not heat-treated. In the metal/HDP prostheses the average wear rate was 0.025 mm/year. The relative wear on the medial and lateral sides was related to the leg axis. None of the retrieved prostheses showed any severe disruption of their surface. The polyacetal/HDP prostheses showed similar wear with a statistically insignificant trend towards slower penetration. We conclude that the rate of wear of HDP in a conforming tibiofemoral bearing with a fixed tibial component at least 6 mm thick and not heat-treated is slow enough to be safe in clinical practice.
引用
收藏
页码:630 / 636
页数:7
相关论文
共 13 条
[1]   POLYETHYLENE WEAR IN MENISCAL KNEE REPLACEMENT - A ONE TO 9-YEAR RETRIEVAL ANALYSIS OF THE OXFORD KNEE [J].
ARGENSON, JN ;
OCONNOR, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02) :228-232
[2]   MECHANICAL TESTS ON TIBIAL COMPONENTS OF NON-HINGED KNEE PROSTHESES [J].
BARGREN, JH ;
DAY, WH ;
FREEMAN, MAR ;
SWANSON, SAV .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (02) :256-261
[3]  
BRADLEY GW, 1992, AM J KNEE SURG, V5, P3
[4]  
Charnley J., 1979, LOW FRICTION ARTHROP
[5]   POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[6]   ICLH ARTHROPLASTY OF KNEE - 1968-1977 [J].
FREEMAN, MAR ;
TODD, RC ;
BAMERT, P ;
DAY, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :339-344
[7]   MECHANICS OF KNEE AND PROSTHESIS DESIGN [J].
GOODFELLOW, J ;
OCONNOR, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :358-369
[8]   POLYETHYLENE WEAR IN UNCEMENTED KNEE REPLACEMENTS [J].
JONES, SMG ;
PINDER, IM ;
MORAN, CG ;
MALCOLM, AJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :18-22
[9]   MASSIVE FEMORAL OSTEOLYSIS AND EARLY DISINTEGRATION OF A POLYETHYLENE-BEARING SURFACE OF A TOTAL KNEE REPLACEMENT - A CASE-REPORT [J].
KILGUS, DJ ;
FUNAHASHI, TT ;
CAMPBELL, PA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (05) :770-774
[10]  
MCKELLOP H, 1985, 31ST P ANN M ORTH RE, P322