FOCUSED EXTRACORPOREAL PYROTHERAPY - FEASIBILITY STUDY IN MAN

被引:98
作者
VALLANCIEN, G [1 ]
HAROUNI, M [1 ]
VEILLON, B [1 ]
MOMBET, A [1 ]
PRAPOTNICH, D [1 ]
BRISSET, JM [1 ]
BOUGARAN, J [1 ]
机构
[1] CTR MED CHIRURG PORTE CHOISY, DEPT PATHOL, PARIS, FRANCE
关键词
D O I
10.1089/end.1992.6.173
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
A feasibility study of focused extracorporeal pyrotherapy was conducted in 50 patients. The instrument has a mobile firing head, which is composed of multiple piezoelectric ceramics with a focal length of 320 mm. The energy focus was 10 mm high x 1.5 mm wide. A retractable and rotating central 3.5-MHz ultrasound transducer was used to localize the tissue target. A 7.5-MHz sectorial transrectal transducer was used for the prostate. The recording of the temperature curves by means of implanted thermocouples demonstrated that the apparatus was able to induce a peak temperature of 120-degrees-C in the focal zone. Heat diffusion was minimal because of the brief shot durations (0.015 to 1 second). Twenty-eight patients with benign prostatic hyperplasia were treated. There was virtually no heat diffusion to the rectum and urethra. The maximum temperature recorded in 12 patients who underwent pyrotherapy for superficial bladder tumors was 65-degrees-C. Eight patients underwent the same treatment to the kidney, and two patients were treated for liver metastases. Histologic examination of the target organs demonstrated severe tissue laceration with intense congestion and marked hyperemia on Day 0 and very early signs of necrosis on Day 2, corresponding to the dimensions of the focal area. No significant alterations in the usual laboratory values were observed apart from a transient rise in creatine phosphokinase with the longer shot durations. The side effects were skin burns. A shot duration-skill focal point distance curve can be used to avoid these adverse effects.
引用
收藏
页码:173 / 181
页数:9
相关论文
共 20 条
[1]  
[Anonymous], 1880, CR HEBD ACAD SCI
[2]  
Chapelon J Y, 1991, Prog Urol, V1, P231
[3]  
FESSENDEN P, 1984, IEEE T ULTRASON FERR, P127
[4]  
FRY F J, 1970, Journal of the Acoustical Society of America, V48, P1413, DOI 10.1121/1.1912301
[5]  
HENLE KJ, 1978, CANCER RES, V38, P1843
[6]   A SCANNED, FOCUSED, MULTIPLE TRANSDUCER ULTRASONIC SYSTEM FOR LOCALIZED HYPERTHERMIA TREATMENTS [J].
HYNYNEN, K ;
ROEMER, R ;
ANHALT, D ;
JOHNSON, C ;
XU, ZX ;
SWINDELL, W ;
CETAS, T .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (01) :21-35
[7]   THE THRESHOLD FOR THERMALLY SIGNIFICANT CAVITATION IN DOG THIGH MUSCLE INVIVO [J].
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (02) :157-169
[8]   THE EFFECT OF BLOOD PERFUSION RATE ON THE TEMPERATURE DISTRIBUTIONS INDUCED BY MULTIPLE, SCANNED AND FOCUSED ULTRASONIC BEAMS IN DOGS KIDNEYS INVIVO [J].
HYNYNEN, K ;
DEYOUNG, D ;
KUNDRAT, M ;
MOROS, E .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1989, 5 (04) :485-497
[9]   TEMPERATURE DISTRIBUTIONS DURING CLINICAL SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA TREATMENTS [J].
HYNYNEN, K ;
SHIMM, D ;
ANHALT, D ;
STEA, B ;
SYKES, H ;
CASSADY, JR ;
ROEMER, RB .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1990, 6 (05) :891-908
[10]   THE CONCENTRIC-RING ARRAY FOR ULTRASOUND HYPERTHERMIA - COMBINED MECHANICAL AND ELECTRICAL SCANNING [J].
IBBINI, MS ;
CAIN, CA .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1990, 6 (02) :401-419