ONLINE CHEMILUMINESCENCE ASSAY USING FIA AND FIBER OPTICS FOR URINARY AND BLOOD-GLUCOSE

被引:9
作者
CATTANEO, MV [1 ]
LUONG, JHT [1 ]
机构
[1] NATL RES COUNCIL CANADA,BIOTECHNOL RES INST,MONTREAL H4P 2R2,PQ,CANADA
关键词
FIBER OPTIC; CHEMILUMINESCENCE; GLUCOSE; ACETATE EXCHANGER;
D O I
10.1016/0141-0229(93)90130-T
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A chemiluminescence fiber optic system coupled to flow injection analysis (FIA) and ion exchange chromatography has been developed for determining glucose in blood and urine. Immobilized glucose oxidase acted on beta-D-glucose to produce hydrogen peroxide, which was then reacted with luminol in the presence of ferricyanide to produce a light signal. Endogenous ascorbic acid and uric acid present in urine or blood samples were effectively retained by an upstream acetate anion exchanger. In addition, acetaminophen could also be adsorbed by this ion exchanger. The detection system exhibited a sensitivity of 1.315 +/- 0.044 RU muM-1 for glucose with a minimum detection level of 1 muM. When applied for the determination of urinary and blood glucose levels, the results obtained compared well with those of the reference hexokinase assay. Immobilized glucose oxidase was reused for over 500 analyses without losing its original activity. A conservative estimate for the reuse of the acetate ion exchange column was about 100 analyses.
引用
收藏
页码:424 / 428
页数:5
相关论文
共 15 条
[1]  
ARNOLD MA, 1991, BIOSENSOR PRINCIPLES, P195
[2]   QUANTITATIVE-DETERMINATION OF BLOOD-GLUCOSE USING ENZYME INDUCED CHEMILUMINESCENCE OF LUMINOL [J].
BOSTICK, DT ;
HERCULES, DM .
ANALYTICAL CHEMISTRY, 1975, 47 (03) :447-452
[3]   FERROCENE-MEDIATED ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE [J].
CASS, AEG ;
DAVIS, G ;
FRANCIS, GD ;
HILL, HAO ;
ASTON, WJ ;
HIGGINS, IJ ;
PLOTKIN, EV ;
SCOTT, LDL ;
TURNER, APF .
ANALYTICAL CHEMISTRY, 1984, 56 (04) :667-671
[4]   A CHEMILUMINESCENCE FIBEROPTIC BIOSENSOR SYSTEM FOR THE DETERMINATION OF GLUTAMINE IN MAMMALIAN-CELL CULTURES [J].
CATTANEO, MV ;
MALE, KB ;
LUONG, JHT .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (08) :569-574
[5]  
CENAS NK, 1984, BIOCHIM BIOPHYS ACTA, V767, P108
[6]  
Clark LC, 1970, US patent, Patent No. [3539455, 3,539,455]
[7]   DIRECT ELECTRICAL COMMUNICATION BETWEEN CHEMICALLY MODIFIED ENZYMES AND METAL-ELECTRODES .1. ELECTRON-TRANSFER FROM GLUCOSE-OXIDASE TO METAL-ELECTRODES VIA ELECTRON RELAYS, BOUND COVALENTLY TO THE ENZYME [J].
DEGANI, Y ;
HELLER, A .
JOURNAL OF PHYSICAL CHEMISTRY, 1987, 91 (06) :1285-1289
[8]   THE POTENTIAL ROLE OF BIOSENSORS IN THE FOOD AND DRINK INDUSTRIES [J].
LUONG, JHT ;
GROOM, CA ;
MALE, KB .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (07) :547-554
[9]  
PETERSON JI, 1967, ANAL BIOCHEM, V23, P301
[10]   ENZYMATIC ASSAY BY FLOW-INJECTION ANALYSIS WITH DETECTION BY CHEMILUMINESCENCE - DETERMINATION OF GLUCOSE, CREATININE, FREE-CHOLESTEROL AND LACTIC-ACID USING AN INTEGRATED FIA MICROCONDUIT [J].
PETERSSON, BA ;
HANSEN, EH ;
RUZICKA, J .
ANALYTICAL LETTERS, 1986, 19 (5-6) :649-665