VARIATIONS IN THE POWER OUTPUT OF THE PIEZON-MASTER-400 ULTRASONIC ENDODONTIC UNIT

被引:5
作者
AHMAD, M
ROY, RA
KAMARUDIN, AG
机构
[1] Department of Conservative Dentistry, University of Malaya, Kuala Lumpur
[2] The Applied Physics Laboratory, The University of Washington, Seattle, Washington
[3] Department of Mechanical Engineering, University of Malaya, Kuala Lumpur
关键词
ENDOSONICS; ROOT CANAL INSTRUMENT; ROOT CANAL THERAPY;
D O I
10.1111/j.1365-2591.1994.tb00225.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The present study was undertaken to see if there was any variability in the power output of Piezon-Master 400 ultrasonic files when driven using different generators, tranducers and file holders. The displacement amplitude of the oscillating tip of the file in air was used as a measure of the power output. The results showed that there was considerable variability in the power output of Piezon-Master 400 ultrasonic files of similar size and length when driven using different generators, transducers and file holders. In consideration of this, it is recommended that a calibration device be incorporated in the ultrasonic unit so that the operator will have some knowledge of when the unit is working at its maximum efficiency.
引用
收藏
页码:26 / 31
页数:6
相关论文
共 12 条
[11]  
Martin H, Ultrasonic disinfection of the root canal, Oral Surgery, Oral Medicine and Oral Pathology, 42, pp. 92-99, (1976)
[12]  
Walmsley AD., Williams AR, Effects of constraint on the oscillatory pattern of endosonic files, Journal of Endodontics, 15, pp. 189-194, (1989)