HARDWARE AND SOFTWARE ARTIFACTS IN STORAGE PHOSPHOR RADIOGRAPHY

被引:37
作者
OESTMANN, JW [1 ]
PROKOP, M [1 ]
SCHAEFER, CM [1 ]
GALANSKI, M [1 ]
机构
[1] HANOVER MED SCH,DEPT DIAGNOST RADIOL,W-3000 HANNOVER 61,GERMANY
关键词
IMAGES; ARTIFACT; QUALITY; RADIOGRAPHY; DIGITAL; STORAGE PHOSPHOR;
D O I
10.1148/radiographics.11.5.1947316
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Hardware and software artifacts in digital radiographs acquired with storage phosphor systems can seriously impair image quality and imitate or mask abnormalities. These artifacts are caused by image plate, image reader, and laser printer defects; faulty image readout; processing errors; and unsharp masking. The artifacts can simulate calcifications and pneumothoraces or conceal low-contrast ill-defined lesions and subtle lesions along opacity interfaces. Hardware artifacts need to be recognized and properly traced to repair the system or improve its maintenance. Artifacts due to software characteristics and image postprocessing must also be identified to ensure adequate system handling and adjustment of postprocessing algorithms.
引用
收藏
页码:795 / 805
页数:11
相关论文
共 5 条
[1]  
Blume H., 1987, Proceedings of the SPIE - The International Society for Optical Engineering, V767, P371
[2]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[3]  
MIYAHARA J, 1987, COMPUTED RADIOGRAPHY, P7
[4]   STORAGE PHOSPHOR VERSUS SCREEN-FILM RADIOGRAPHY - EFFECT OF VARYING EXPOSURE PARAMETERS AND UNSHARP MASK FILTERING ON THE DETECTABILITY OF CORTICAL BONE DEFECTS [J].
PROKOP, M ;
GALANSKI, M ;
OESTMANN, JW ;
VONFALKENHAUSEN, U ;
ROSENTHAL, H ;
REIMER, P ;
NISCHELSKY, J ;
REICHELT, S .
RADIOLOGY, 1990, 177 (01) :109-113
[5]   COMPUTED RADIOGRAPHY UTILIZING SCANNING LASER STIMULATED LUMINESCENCE [J].
SONODA, M ;
TAKANO, M ;
MIYAHARA, J ;
KATO, H .
RADIOLOGY, 1983, 148 (03) :833-838