We have examined the effect of screen-structure, optical-detector, and secondary-quantum noise sources on detective quantum efficiency, DQE(f). This was done by using experimental measurements of screen-structure and optical-detector noise in combination with a theoretical model which predicts x-ray quantum and secondary-quantum noise for different optical and physical properties of a phosphor screen. The reduction in DQE(f) from noise sources other than x-ray quantum noise depends on the noise power spectra (NPS) of these other sources relative to the x-ray quantum NPS. Even though x-ray quantum noise may be the dominant noise source at low spatial frequencies, it decreases relatively rapidly with increasing frequency so that other noise sources, which may be small at low frequencies, dominate. Our model predicts that DQE(f) can be increased, at spatial frequencies < 4 mm-1by changing the optical properties of the screen even though modulation transfer function MTF(f) may decrease. Furthermore, if screen and optical-detector noise decrease with increasing frequency and secondary-quantum noise sufficiently small, then DQE(f) will also be improved at frequencies > 4 mm-1. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:887 / 893
页数:7
相关论文
共 22 条