EVALUATION AND TEST OF THE RESPONSE MATRIX OF A MULTISPHERE NEUTRON SPECTROMETER IN A WIDE ENERGY-RANGE .2. SIMULATION

被引:11
作者
AROUA, A [1 ]
GRECESCU, M [1 ]
LANFRANCHI, M [1 ]
LERCH, P [1 ]
PRETRE, S [1 ]
VALLEY, JF [1 ]
机构
[1] SWISS NUCL SAFETY INSPECTORATE,CH-5232 VILLIGEN,SWITZERLAND
关键词
D O I
10.1016/0168-9002(92)90405-S
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The response matrix of a multisphere neutron spectrometer has been calculated using the ANISN particles transport code and a recent condensed cross-section library. The matrix has then been extended up to 400 MeV. A detailed investigation of several calculation parameters led to the establishment of an improved model of the system. The adjustment of the calculated response functions to the experimental calibration points (8 keV-14.8 MeV) has given coherent results, the deviation of the calculated values from the measured ones being less than 10% for the maximum sensitivity regions. The calculated response functions have been compared to others reported in the literature. Finally the effect of a cadmium shell on the spheres' response has been studied.
引用
收藏
页码:305 / 311
页数:7
相关论文
共 38 条
[1]  
ALEVRA A, 1990, COMMUNICATION
[2]  
ALSMILLER RG, 1986, ORNLTM801
[3]   EVALUATION AND TEST OF THE RESPONSE MATRIX OF A MULTISPHERE NEUTRON SPECTROMETER IN A WIDE ENERGY-RANGE .1. CALIBRATION [J].
AROUA, A ;
GRECESCU, M ;
LERCH, P ;
PRETRE, S ;
VALLEY, JF ;
VYLET, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 321 (1-2) :298-304
[4]   EVALUATION AND TEST OF THE RESPONSE MATRIX OF A MULTISPHERE NEUTRON SPECTROMETER IN A WIDE ENERGY-RANGE .3. VALIDATION [J].
AROUA, A ;
GRECESCU, M ;
LERCH, P ;
PRETRE, S ;
VALLEY, JF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 321 (1-2) :312-316
[5]   APPLICATIONS OF BONNER SPHERE DETECTORS IN NEUTRON FIELD DOSIMETRY [J].
AWSCHALOM, M ;
SANNA, RS .
RADIATION PROTECTION DOSIMETRY, 1985, 10 (1-4) :89-101
[6]   A NEW TYPE OF NEUTRON SPECTROMETER [J].
BRAMBLETT, RL ;
EWING, RI ;
BONNER, TW .
NUCLEAR INSTRUMENTS & METHODS, 1960, 9 (01) :1-12
[7]  
CARLSON BG, 1968, REACTOR PHYSICS
[8]  
CAZIERGUES R, 1972, CEAR4400
[9]  
DEVILLERS C, 1970, CEAN1358 NOT
[10]  
DHAIRYAWAN MP, 1980, NUCL INSTRUM METHODS, V169, P15