SCANNING TUNNELING MICROSCOPE CALIBRATION AND RECONSTRUCTION OF REAL IMAGE - DRIFT AND SLOPE ELIMINATION

被引:68
作者
YUROV, VY
KLIMOV, AN
机构
[1] General Physics Institute, Academy of Sciences of Russia, 117942 Moscow
关键词
D O I
10.1063/1.1144890
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Drift, slope of sample, and indeterminate sensitivity of piezoceramics are considered as the origin of the linear scanning tunneling microscope (STM) image distortions. A special algorithm of STM scanning and reconstruction of image of unknown atomic structures is suggested and tested. This algorithm allows us to measure three components of the drift velocity and two angles, characterizing the average slope of scanning surface. On the one hand, using this algorithm, one can perform the STM calibration by a known surface structure (for example, the highly oriented pyrolytic graphite surface) even in the presence of a drift. This enables us to determine all the three piezoceramics constants of STM piezoscanner and the deviation of the real scanner axes X and Y from orthogonality. On the other hand, using such a calibrated STM and the described algorithm, it is possible to obtain the real STM image and make measurements for unknown surfaces with atomic resolution without the distortions mentioned above. As it has been proved, the error of interatomic distance measurements does not exceed 5% (approximately 0.1 angstrom) for any direction along the surface, even in the presence of a drift up to 0.4 angstrom/s and the sample slope angle up to 10-degrees.
引用
收藏
页码:1551 / 1557
页数:7
相关论文
共 12 条
[1]   COMPARISON OF VARIOUS LOW-TEMPERATURE SCANNING TUNNELING MICROSCOPY DESIGNS FOR USE IN SPECTROSCOPY AND TOPOGRAPHY [J].
ADLER, JG ;
CHEN, TT ;
GALLAGHER, MC ;
KONKIN, MK ;
MULLIN, DP .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02) :992-995
[2]   SCANNING TUNNELING MICROSCOPY OF SILVER, GOLD, AND ALUMINUM MONOMERS AND SMALL CLUSTERS ON GRAPHITE [J].
GANZ, E ;
SATTLER, K ;
CLARKE, J .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1988, 6 (02) :419-423
[3]   SCANNING TUNNELING MICROSCOPY OF INDIVIDUAL MOLECULES OF COPPER PHTHALOCYANINE ADSORBED ON POLYCRYSTALLINE SILVER SURFACES [J].
GIMZEWSKI, JK ;
STOLL, E ;
SCHLITTLER, RR .
SURFACE SCIENCE, 1987, 181 (1-2) :267-277
[4]   INITIAL-STAGE OF AU ADSORPTION ONTO A SI(111) SURFACE STUDIED BY SCANNING TUNNELING MICROSCOPY [J].
HASEGAWA, T ;
TAKATA, K ;
HOSAKA, S ;
HOSOKI, S .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02) :758-760
[5]   CRYSTALLINE LATTICE FOR METROLOGICAL APPLICATIONS AND POSITIONING CONTROL BY A DUAL TUNNELING-UNIT SCANNING TUNNELING MICROSCOPE [J].
KAWAKATSU, H ;
HOSHI, Y ;
HIGUCHI, T ;
KITANO, H .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02) :651-654
[6]   DEFORMATIONS AND NONLINEARITY IN SCANNING TUNNELING MICROSCOPE IMAGES [J].
LIBIOULLE, L ;
RONDA, A ;
TABORELLI, M ;
GILLES, JM .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02) :655-658
[7]   SOME DESIGN CRITERIA IN SCANNING TUNNELING MICROSCOPY [J].
POHL, DW .
IBM JOURNAL OF RESEARCH AND DEVELOPMENT, 1986, 30 (04) :417-427
[8]   DIFFRACTION GRATING CALIBRATION OF SCANNING TUNNELING MICROSCOPE PIEZOSCANNERS [J].
POIRIER, GE ;
WHITE, JM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (12) :3917-3918
[9]   THERMAL-EXPANSION AND PIEZOELECTRIC RESPONSE OF PZT CHANNEL-5800 FOR USE IN LOW-TEMPERATURE SCANNING TUNNELING MICROSCOPE DESIGNS [J].
SIMPSON, AM ;
WOLFS, W .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (11) :2193-2195
[10]   RESTORATION OF STM IMAGES DISTORTED BY TIME-DEPENDENT PIEZO DRIVER AFTEREFFECTS [J].
STOLL, EP .
ULTRAMICROSCOPY, 1992, 42 :1585-1589