RISE IN CHRONIC DEFIBRILLATION THRESHOLDS IN NONTHORACOTOMY IMPLANTABLE DEFIBRILLATOR

被引:65
作者
VENDITTI, FJ
MARTIN, DT
VASSOLAS, G
BOWEN, S
机构
[1] Cardiac Electrophysiology Laboratory, Section of Cardiology, Lahey Clinic Medical Center, Burlington, MA
[2] Cardiology Section, Lahey Clinic Medical Center, Burlington, MA 01805
关键词
DEFIBRILLATION; ARRHYTHMIA;
D O I
10.1161/01.CIR.89.1.216
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background To establish the chronic stability of defibrillation thresholds (DFTs) in a transvenous cardioverter/defibrillator (TCD) system, we studied 37 consecutive patients with TCD systems implanted for >6 months. Methods and Results DFT was measured by a step-down method at implant and 2 and 6 months later. The mean ejection fraction was 34.5+/-14.3%. Coronary artery disease with previous myocardial infarction was present in 31 patients. The mean DFT rose from 13.3+/-4.3 J at implant to 16.5+/-4.7 J at 2 months (P<.001) and 17.6+/-5.4 J at 6 months (P<.0001). ANOVA revealed a statistically significant rise in DFT over time (P<.0005). At 2 months, 25 patients had a rise in DFT, and 14 had a rise greater than or equal to 5 J. The observed rise at 2 months persisted in 19 patients. A chronic rise, defined as greater than or equal to 5 J rise at 6 months, occurred in 17 patients. Univariate analysis of clinical as well as implant variables revealed no predictors of a rise in DFT in this group, Conclusions We conclude that there is a significant rise in DFT at 2 and 6 months in this TCD system. Although the chronic threshold remained well within the available energy range of the pulse generator, this observation has important implications for implantation guidelines, programming, and future pulse generator development for TCD patients.
引用
收藏
页码:216 / 223
页数:8
相关论文
共 38 条
[11]   THE DEFIBRILLATION SUCCESS RATE VERSUS ENERGY RELATIONSHIP .1. CURVE FITTING AND THE MOST EFFICIENT DEFIBRILLATION ENERGY [J].
GLINER, BE ;
MURAKAWA, Y ;
THAKOR, NV .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (03) :326-338
[12]  
HALPERIN BD, 1993, PACE PACING CLIN ELE, V16, P854
[13]   CLINICAL-RESULTS WITH NONTHORACOTOMY ICD SYSTEMS [J].
HAUSER, RG ;
KURSCHINSKI, DT ;
MCVEIGH, K ;
THOMAS, A ;
MOWER, MM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01) :141-148
[14]  
HSIA HH, 1992, CIRCULATION S1, V86, pI451
[15]   EFFECTS OF LONG-TERM AMIODARONE THERAPY ON THE DEFIBRILLATION THRESHOLD AND THE RATE OF SHOCKS OF THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
HUANG, SKS ;
DEGUZMAN, WLT ;
CHENARIDES, JG ;
OKIKE, O ;
VANDERSALM, TJ .
AMERICAN HEART JOURNAL, 1991, 122 (03) :720-727
[16]   CURRENT CONCEPTS FOR SELECTING THE LOCATION, SIZE AND SHAPE OF DEFIBRILLATION ELECTRODES [J].
IDEKER, RE ;
WOLF, PD ;
ALFERNESS, C ;
KRASSOWSKA, W ;
SMITH, WM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (02) :227-&
[17]   DEFIBRILLATION EFFICACY - COMPARISON OF DEFIBRILLATION THRESHOLD VERSUS DOSE-RESPONSE CURVE DETERMINATION [J].
JONES, DL ;
IRISH, WD ;
KLEIN, GJ .
CIRCULATION RESEARCH, 1991, 69 (01) :45-51
[18]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS AND SURVIVAL IN PATIENTS WITH MALIGNANT VENTRICULAR ARRHYTHMIAS [J].
KELLY, PA ;
CANNOM, DS ;
GARAN, H ;
MIRABAL, GS ;
HARTHORNE, JW ;
HURVITZ, RJ ;
VLAHAKES, GJ ;
JACOBS, ML ;
ILVENTO, JP ;
BUCKLEY, MJ ;
RUSKIN, JN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) :1278-1286
[19]   AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR IMPLANTATION WITHOUT THORACOTOMY USING AN ENDOCARDIAL AND SUBMUSCULAR PATCH SYSTEM [J].
MCCOWAN, R ;
MALONEY, J ;
WILKOFF, B ;
SIMMONS, T ;
KHOURY, D ;
MCALISTER, H ;
MORANT, V ;
CASTLE, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (02) :415-421
[20]  
MCDANIEL WC, 1987, MED INSTRUM, V21, P170