DETECTION OF ERRORS IN INDIVIDUAL PATIENTS IN RADIOTHERAPY BY SYSTEMATIC IN-VIVO DOSIMETRY

被引:78
作者
NOEL, A [1 ]
ALETTI, P [1 ]
BEY, P [1 ]
MALISSARD, L [1 ]
机构
[1] CTR ALEXIS VAUTRIN,SERV RADIOTHERAPY,F-54511 VANDOEUVRE NANCY,FRANCE
关键词
IN VIVO DOSIMETRY; PLANNING ERRORS; TREATMENT ERRORS; RADIOTHERAPY; SILICON DIODES; QUALITY ASSURANCE;
D O I
10.1016/0167-8140(94)01503-U
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
We report 5 years of systematic measurements of the dose delivered to each patient undergoing radiotherapy treatment with photon beams in order to detect any systematic error that may have escaped the different checks performed at each step of planning and calculation prior to the first treatment session, or may have arisen during the set-up or the treatment delivery. For each patient the target-absorbed dose is derived from the entrance and exit doses measured by silicon diodes, on the beam axis at the patient's skin. Depending on the discrepancies observed between the measured and expected doses we have set decision levels for the corrective actions to be taken. In addition these measurements allow us to obtain information on the overall accuracy or on the quality of a specific treatment. During 5 years, 7519 patients have been measured and 79 errors were detected. Half could have induced a variation of over 10% in the dose delivered. Seventy-eight out of 79 errors were of human origin. As part of an overall quality assurance programme, it is of the utmost importance to check the dose delivered for each patient undergoing radiotherapy treatment in order to avoid systematic underdosing or overdosing.
引用
收藏
页码:144 / 151
页数:8
相关论文
共 23 条
[1]  
ALETTI P, 1991, B CANCER RADIOTHER, V78, P535
[2]   HUMAN ERRORS IN THE CALCULATION OF MONITOR UNITS IN CLINICAL RADIOTHERAPY PRACTICE [J].
CALANDRINO, R ;
CATTANEO, GM ;
DELVECCHIO, A ;
FIORINO, C ;
LONGOBARDI, B ;
SIGNOROTTO, P .
RADIOTHERAPY AND ONCOLOGY, 1993, 28 (01) :86-88
[3]   THE ROLE OF IN-VIVO DOSIMETRY [J].
GARAVAGLIA, G ;
JOHANSSON, KA ;
LEUNENS, G ;
MIJNHEER, B .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :281-282
[4]  
HERRING DF, 1971, COMPUTERS RADIOTHERA, P51
[5]   COMPARISON OF ENTRANCE AND EXIT DOSE MEASUREMENTS USING IONIZATION CHAMBERS AND SILICON DIODES [J].
HEUKELOM, S ;
LANSON, JH ;
MIJNHEER, BJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (01) :47-59
[6]   INVIVO DOSIMETRY DURING TANGENTIAL BREAST TREATMENT [J].
HEUKELOM, S ;
LANSON, JH ;
VANTIENHOVEN, G ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) :269-279
[7]   INVIVO DOSIMETRY DURING PELVIC TREATMENT [J].
HEUKELOM, S ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1992, 25 (02) :111-120
[8]   QUALITY ASSURANCE IN RADIOTHERAPY BY INVIVO DOSIMETRY .1. ENTRANCE DOSE MEASUREMENTS, A RELIABLE PROCEDURE [J].
LEUNENS, G ;
VANDAM, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1990, 17 (02) :141-151
[9]   QUALITY ASSURANCE IN RADIOTHERAPY BY INVIVO DOSIMETRY .2. DETERMINATION OF THE TARGET ABSORBED DOSE [J].
LEUNENS, G ;
VANDAM, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1990, 19 (01) :73-87
[10]   HUMAN ERRORS IN DATA TRANSFER DURING THE PREPARATION AND DELIVERY OF RADIATION TREATMENT AFFECTING THE FINAL RESULT - GARBAGE IN, GARBAGE OUT [J].
LEUNENS, G ;
VERSTRAETE, J ;
VANDENBOGAERT, W ;
VANDAM, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1992, 23 (04) :217-222