FLOW-INJECTION SYSTEMS FOR DIRECTLY COUPLING ONLINE DIGESTIONS WITH ANALYTICAL ATOMIC SPECTROMETRY .1. DISSOLUTION OF COCOA UNDER STOPPED-FLOW, HIGH-PRESSURE CONDITIONS

被引:16
作者
GLUODENIS, TJ [1 ]
TYSON, JF [1 ]
机构
[1] UNIV MASSACHUSETTS,DEPT CHEM,AMHERST,MA 01003
关键词
FLOW INJECTION ATOMIC ABSORPTION SPECTROMETRY; SLURRY SAMPLING; ONLINE DIGESTION; COPPER AND IRON DETERMINATION; COCOA ANALYSIS;
D O I
10.1039/ja9920700301
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A double flow injection manifold incorporating a resistively heated oven is under development for the direct coupling of the digestion of solid samples with an analytical spectrometric technique. The potential of the system was illustrated by the dissolution of cocoa powder and subsequent determination of the copper and iron content by flame atomic absorption spectrometry (FAAS). The cocoa powder was slurried in 10% nitric acid, injected into the manifold and digested under stopped-flow, high-pressure conditions. Gas-liquid separation was effected by a two-stage de-pressurization system. Copper and iron were determined by flow injection FAAS against acid matched standards. Values of 44 +/- 19 and 144 +/- 10 mg kg-1 were obtained for the copper and iron content, respectively. The cocoa powder was also digested using an open-vessel hot-plate method and a closed-vessel microwave digestion method. The results for the copper content were 52 +/- 2 and 50 +/- 2 mg kg-1, respectively. Both of these methods gave a value of 180 +/- 10 mg kg-1 for iron. The relatively large uncertainty in the copper result for the on-line method was owing to the higher dilution in this system producing a concentration in the digest nearer the detection limit of the flame procedure. The low results for iron were considered to be owing to incomplete digestion, as agreement with the results of the other methods was obtained if the solution was analysed by the standard additions method. The proposed procedure produces a clear solution, and allows increased sample throughput while minimizing sample contamination and decreasing sample and reagent consumption.
引用
收藏
页码:301 / 306
页数:6
相关论文
共 21 条
[1]  
[Anonymous], 1988, INTRO MICROWAVE SAMP
[2]  
APPLETON JMH, 1986, ANAL CHIM ACTA, V179, P269, DOI 10.1016/S0003-2670(00)84471-2
[3]   DETERMINATION OF ZINC AND CADMIUM IN SMALL AMOUNTS OF BIOLOGICAL TISSUES BY MICROWAVE-ASSISTED DIGESTION AND FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY [J].
BURGUERA, M ;
BURGUERA, JL ;
ALARCON, OM .
ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) :421-427
[4]   FLOW-INJECTION AND MICROWAVE-OVEN SAMPLE DECOMPOSITION FOR DETERMINATION OF COPPER, ZINC AND IRON IN WHOLE-BLOOD BY ATOMIC-ABSORPTION SPECTROMETRY [J].
BURGUERA, M ;
BURGUERA, JL ;
ALARCON, OM .
ANALYTICA CHIMICA ACTA, 1986, 179 :351-357
[5]   ONLINE MICROWAVE-OVEN DIGESTION FLAME ATOMIC-ABSORPTION ANALYSIS OF SOLID SAMPLES [J].
CARBONELL, V ;
DELAGUARDIA, M ;
SALVADOR, A ;
BURGUERA, JL ;
BURGUERA, M .
ANALYTICA CHIMICA ACTA, 1990, 238 (02) :417-421
[6]   A STUDY OF LASER ABLATION AND SLURRY NEBULIZATION SAMPLE INTRODUCTION FOR THE ANALYSIS OF GEOCHEMICAL MATERIALS BY INDUCTIVELY COUPLED PLASMA SPECTROMETRY [J].
DARKE, SA ;
LONG, SE ;
PICKFORD, CJ ;
TYSON, JF .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (03) :284-289
[7]   LASER ABLATION SYSTEM FOR SOLID SAMPLE ANALYSIS BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
DARKE, SA ;
LONG, SE ;
PICKFORD, CJ ;
TYSON, JF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :715-719
[8]  
DELAGUARDIA M, 1988, J FLOW INJECTION ANA, V5, P121
[10]   GENERAL-ASPECTS OF TRACE ANALYTICAL METHODS .6. ACID PRESSURE DECOMPOSITION IN TRACE-ELEMENT ANALYSIS [J].
JACKWERTH, E ;
GOMISCEK, S .
PURE AND APPLIED CHEMISTRY, 1984, 56 (04) :479-489