AUTOMATED BIOPSY DEVICES - A BLINDED EVALUATION

被引:67
作者
HOPPER, KD
ABENDROTH, CS
STURTZ, KW
MATTHEWS, YL
STEVENS, LA
SHIRK, SJ
机构
[1] PENN STATE UNIV,MILTON S HERSHEY MED CTR,DEPT PATHOL,HERSHEY,PA 17033
[2] PENN STATE UNIV,MILTON S HERSHEY MED CTR,CTR BIOSTAT & EPIDEMIOL,HERSHEY,PA 17033
关键词
BIOPSIES; TECHNOLOGY; KIDNEY; BIOPSY; LIVER; MUSCLES; PSOAS; PANCREAS; PROSTATE; UTERUS;
D O I
10.1148/radiology.187.3.8497611
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate 20 different automated biopsy devices with respect to the quality of tissue obtained for histopathologic analysis, a total of 1,470 18-gauge biopsy specimens were obtained from 10 fresh autopsy cases, including 30 liver, 20 kidney, 10 pancreas, and 10 psoas muscle biopsy specimens per device and per biopsy depth. There was no statistical difference in the performance of the long-throw Biopty, ASAP 18, 1.9-cm Ultra-Cut, long-throw Monopty, and 2.5-cm ABS biopsy guns. All obtained a large amount of tissue with minimal fragmentation or crush artifact. Most of the short-throw biopsy guns (depth of biopsy less-than-or-equal-to 1.1 cm) did not perform as well. Although the other guns performed adequately, less than optimal results were obtained with the Temno, Bio-Gun, Roth, Klear Kut, ABC, and Urocut biopsy guns. Most 18-gauge automated biopsy devices with a biopsy excursion of at least 2.0 cm provide a high-quality, diagnostically adequate specimen for histopathologic analysis.
引用
收藏
页码:653 / 660
页数:8
相关论文
共 19 条
  • [1] PANCREATIC TRANSPLANTS - CT-GUIDED BIOPSY
    BERNARDINO, M
    FERNANDEZ, M
    NEYLAN, J
    HERTZLER, G
    WHELCHEL, J
    OLSON, R
    [J]. RADIOLOGY, 1990, 177 (03) : 709 - 711
  • [2] AUTOMATED BIOPSY DEVICES - SIGNIFICANCE AND SAFETY
    BERNARDINO, ME
    [J]. RADIOLOGY, 1990, 176 (03) : 615 - 616
  • [3] LIVER-TRANSPLANT BIOPSIES WITH A BIOPSY GUN
    CHEZMAR, JL
    KEITH, LL
    NELSON, RC
    PLAIRE, JC
    HERTZLER, GL
    BERNARDINO, ME
    [J]. RADIOLOGY, 1991, 179 (02) : 447 - 448
  • [4] ULTRASOUND GUIDED CORE BIOPSY IN AIDS - EXPERIENCE IN 6 PATIENTS
    DONALD, JJ
    CORAL, A
    SHORVON, PJ
    LEES, WR
    [J]. BRITISH MEDICAL JOURNAL, 1988, 296 (6622) : 606 - 607
  • [5] BIOPSY OF THE PANCREAS WITH A BIOPSY GUN
    ELVIN, A
    ANDERSSON, T
    SCHEIBENPFLUG, L
    LINDGREN, PG
    [J]. RADIOLOGY, 1990, 176 (03) : 677 - 679
  • [6] FROEDE RC, 1990, HDB FORENSIC PATHYOL, P71
  • [7] EFFICACY OF AUTOMATED BIOPSY GUNS VERSUS CONVENTIONAL BIOPSY NEEDLES IN THE PYGMY PIG
    HOPPER, KD
    BAIRD, DE
    REDDY, VV
    LANDIS, JR
    PARKER, SH
    TYLER, HN
    OWNBEY, JL
    MCCAUSLIN, MA
    YAKES, WF
    SABATELLI, FW
    MOTT, PL
    [J]. RADIOLOGY, 1990, 176 (03) : 671 - 676
  • [8] FINE-NEEDLE ASPIRATION BIOPSY FOR CYTOPATHOLOGIC ANALYSIS - UTILITY OF SYRINGE HANDLES, AUTOMATED GUNS, AND THE NONSUCTION METHOD
    HOPPER, KD
    ABENDROTH, CS
    STURTZ, KW
    MATTHEWS, YL
    SHIRK, SJ
    [J]. RADIOLOGY, 1992, 185 (03) : 819 - 824
  • [9] PERCUTANEOUS NEEDLE-BIOPSY - A NEW TECHNIQUE
    LINDGREN, PG
    [J]. ACTA RADIOLOGICA-DIAGNOSIS, 1982, 23 (06): : 653 - 656
  • [10] Mladinich C. R. J., 1991, AALAS Bulletin, V30, P9