PERFORMANCE EVALUATION OF THE POSITRON SCANNER ECAT EXACT

被引:211
作者
WIENHARD, K
ERIKSSON, L
GROOTOONK, S
CASEY, M
PIETRZYK, U
HEISS, WD
机构
[1] MAX PLANCK INST NEUROL RES,GLEUELER STR 50,W-5000 COLOGNE 41,GERMANY
[2] KAROLINSKA HOSP,S-10401 STOCKHOLM 60,SWEDEN
[3] CTI INC,KNOXVILLE,TN
[4] HAMMERSMITH HOSP,LONDON W12 0HS,ENGLAND
关键词
EMISSION COMPUTED TOMOGRAPHY; APPARATUS AND EQUIPMENT; PERFORMANCE EVALUATION; TECHNIQUES;
D O I
10.1097/00004728-199209000-00024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Cologne Special is a prototype of the ECAT EXACT (mode 921), a new generation of Siemens-CTI PET scanners. It consists of three rings of 48 BGO block detectors each, covering an axial field of view of 16.2 cm with a patient port of 56.2 cm diameter. This results in a total of 24 rings with 384 crystals each, giving 47 contiguous image planes in two-dimensional (2D) mode. Total system sensitivity is 216 kcps/mu-Ci/ml for a 20 cm cylinder phantom in 2D. This increases to 1.5 Mcps/mi-Ci/ml in 3D. Data are acquired in the stationary mode only (no wobble motion), resulting in a transaxial spatial resolution of better than 6 mm full width at half-maximum (FWHM) at the center, which degrades to 7.5 mm tangentially and 9.6 mm radially at a radius of 20 cm. Average axial resolution changes from 5.0 mm FWHM at the center to 8.1 mm at R = 20 cm. Count rate performance was investigated at different low energy discriminator settings and found to be linear up to 2.5-mu-Ci/ml with a 20 cm phantom. The magnitude and distribution of scatter were evaluated for both septa-extended and septa-retracted conditions for a range of energy thresholds. Brain, heart, and whole-body studies with the new tomograph demonstrate the versatility of its applications without compromising on physical performance.
引用
收藏
页码:804 / 813
页数:10
相关论文
共 12 条
  • [1] CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS
    BERGSTROM, M
    ERIKSSON, L
    BOHM, C
    BLOMQVIST, G
    LITTON, J
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) : 42 - 50
  • [2] A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY
    CASEY, ME
    NUTT, R
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) : 460 - 463
  • [3] HIGH-RESOLUTION COMPUTED TOMOGRAPHY OF POSITRON EMITTERS
    DERENZO, SE
    BUDINGER, TF
    CAHOON, JL
    HUESMAN, RH
    JACKSON, HG
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) : 544 - 558
  • [4] FIGURES OF MERIT FOR DIFFERENT DETECTOR CONFIGURATIONS UTILIZED IN HIGH-RESOLUTION POSITRON CAMERAS
    ERIKSSON, L
    BERGSTROM, M
    BOHM, C
    HOLTE, S
    KESSELBERG, M
    LITTON, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) : 446 - 451
  • [5] A 4 RING POSITRON CAMERA SYSTEM FOR EMISSION TOMOGRAPHY OF THE BRAIN
    ERIKSSON, L
    BOHM, C
    KESSELBERG, M
    BLOMQVIST, G
    LITTON, J
    WIDEN, L
    BERGSTROM, M
    ERICSON, K
    GREITZ, T
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1982, 29 (01) : 539 - 543
  • [6] A VMEBUS BASED, REAL-TIME SORTER DESIGN FOR POSITRON EMISSION TOMOGRAPHY
    JONES, WF
    CASEY, ME
    BYARS, LG
    BURGISS, SG
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) : 601 - 604
  • [7] PHYSICAL CHARACTERISTICS OF THE ECAT 953B/31 - A NEW HIGH-RESOLUTION BRAIN POSITRON TOMOGRAPH
    MAZOYER, B
    TREBOSSEN, R
    DEUTCH, R
    CASEY, M
    BLOHM, K
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (04) : 499 - 504
  • [8] MOYERS JC, 1990, THESIS U TENNESSEE
  • [9] COINCIDENCE DETECTION AND SELECTION IN POSITRON EMISSION TOMOGRAPHY USING VLSI
    NEWPORT, DF
    DENT, HM
    CASEY, ME
    BOULDIN, DW
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) : 1052 - 1055
  • [10] PHYSICAL PERFORMANCE OF THE LATEST GENERATION OF COMMERCIAL POSITRON SCANNER
    SPINKS, TJ
    JONES, T
    GILARDI, MC
    HEATHER, JD
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) : 721 - 725