ACCURACY OF FIELD ALIGNMENT IN RADIOTHERAPY OF HEAD AND NECK-CANCER UTILIZING INDIVIDUALIZED FACE MASK IMMOBILIZATION - A RETROSPECTIVE ANALYSIS OF CLINICAL-PRACTICE

被引:54
作者
HESS, CF [1 ]
KORTMANN, RD [1 ]
JANY, R [1 ]
HAMBERGER, A [1 ]
BAMBERG, M [1 ]
机构
[1] UNIV TUBINGEN, DEPT RADIOTHERAPY, TUBINGEN, GERMANY
关键词
RADIOTHERAPY; HEAD AND NECK CANCER; ACCURACY; REPRODUCIBILITY;
D O I
10.1016/0167-8140(94)01497-Q
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Deviations between simulation and first check films were quantitatively assessed for 95 unselected head and neck cancer patients. All measured deviations - calculated on the basis of a total of 190 simulation and 380 verification films - were normally distributed, with mean values of 0-3 mm and standard deviations of 3-5 mm. Of the absolute deviations, 50% and 95% were within 3 mm and 9 mm, respectively. These results should be considered in clinical practice when prescribing safety margins and adequate cut off doses for sparing critical organs in head and neck cancer.
引用
收藏
页码:69 / 72
页数:4
相关论文
共 22 条
[1]  
BEL M, 1993, RRADIOTHER ONCOL, V29, P253
[2]  
BLANCO S, 1987, Radiotherapy and Oncology, V8, P256
[3]   WEEKLY LOCALIZATION FILMS AND DETECTION OF FIELD PLACEMENT ERRORS [J].
BYHARDT, RW ;
COX, JD ;
HORNBURG, A ;
LIERMANN, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1978, 4 (9-10) :881-887
[4]   OBJECTIVE DECISION-MAKING FOLLOWING A PORTAL FILM - THE RESULTS OF A PILOT-STUDY [J].
DENHAM, JW ;
DALLY, MJ ;
HUNTER, K ;
WHEAT, J ;
FAHEY, PP ;
HAMILTON, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (05) :869-876
[5]   QUALITY-CONTROL AT THE PATIENT LEVEL - ACTION OR RETROSPECTIVE INTROSPECTION [J].
DUTREIX, A ;
VANDERSCHUEREN, E ;
LEUNENS, L .
RADIOTHERAPY AND ONCOLOGY, 1992, 25 (02) :146-147
[6]   IMMOBILIZATION ERROR - SOME THEORETICAL CONSIDERATIONS [J].
GOITEIN, M ;
BUSSE, J .
RADIOLOGY, 1975, 117 (02) :407-412
[7]   QUALITY-CONTROL IN RADIOTHERAPY - THE REDUCTION OF FIELD PLACEMENT ERRORS [J].
GRIFFITHS, SE ;
PEARCEY, RG ;
THOROGOOD, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (10) :1583-1588
[8]  
Hess C. F., 1993, Strahlentherapie und Onkologie, V169, P660
[9]   ACCURACY IN RADIATION-FIELD ALIGNMENT IN HEAD AND NECK-CANCER - A PROSPECTIVE-STUDY [J].
HUIZENGA, H ;
LEVENDAG, PC ;
DEPORRE, PMZR ;
VISSER, AG .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (02) :181-187
[10]   LOCALIZATION ERRORS IN MANTLE-FIELD IRRADIATION FOR HODGKINS-DISEASE [J].
HULSHOF, M ;
VANUYTSEL, L ;
VANDENBOGAERT, W ;
VANDERSCHUEREN, E .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (03) :679-683