EVALUATION OF AUTOMATED RETICULOCYTE COUNTS AND THEIR RELIABILITY IN THE PRESENCE OF HOWELL-JOLLY BODIES

被引:20
作者
LOFSNESS, KG
KOHNKE, ML
GEIER, NA
机构
[1] University of Minnesota, Box 198 UMHC, Minneapolis
关键词
AUTOMATED COUNTING; FLOW CYTOMETRY; HOWELL-JOLLY BODIES; RETICULOCYTES;
D O I
10.1093/ajcp/101.1.85
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
An automated reticulocyte procedure using a flow cytometer and the fluorescent dye thiazole orange was evaluated for clinical use. The mean reticulocyte count on 118 hematologically healthy adults was 1.56% (standard deviation [SD] 0.54), with virtually no difference in percentage between women and men. The mean absolute values were 68.4 x 10(9)/L (SD 24.6) and 75.7 X 10(9)/L (SD 27.2), respectively. When compared with the standard microscopic technique, the automated method showed excellent correlation (r = a.98) and greatly improved precision (coefficient of variation [CV] 4.1%) over the manual method (CV 22.8%). Preanalytic storage of blood samples at 4 degrees C for up to 48 hours did not significantly affect results, nor did varying the incubation time of diluted samples from 1/2 to 2 hours. In a group of patients with appreciable numbers of Howell-Jolly bodies, automated reticulocyte counts were spuriously elevated. The difference between the manual and automated counts on these patients approximated the percentage of Howell-Jolly bodies observed on their Wright-Giemsa stained blood smears.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 19 条
[1]  
BATJER JD, 1991, HEMATOLOGY MONOGRAPH, P1
[2]   COMPARISON OF A MODIFIED THIAZOLE ORANGE TECHNIQUE WITH A FULLY AUTOMATED ANALYZER FOR RETICULOCYTE COUNTING [J].
BOWEN, D ;
BENTLEY, N ;
HOY, T ;
CAVILL, I .
JOURNAL OF CLINICAL PATHOLOGY, 1991, 44 (02) :130-133
[3]  
BULL BS, 1990, HEMATOLOGY, P308
[4]   COUNTING RETICULOCYTES BY FLOW-CYTOMETRY - USE OF THIAZOLE ORANGE [J].
CARTER, JM ;
MCSWEENEY, PA ;
WAKEM, PJ ;
NEMET, AM .
CLINICAL AND LABORATORY HAEMATOLOGY, 1989, 11 (03) :267-271
[5]   ENUMERATION OF RETICULOCYTES USING FLUORESCENCE-ACTIVATED FLOW-CYTOMETRY [J].
CORASH, L ;
RHEINSCHMIDT, M ;
LIEU, S ;
MEERS, P ;
BREW, E .
PATHOLOGY AND IMMUNOPATHOLOGY RESEARCH, 1988, 7 (05) :381-394
[6]  
DAVIS BH, 1989, ARCH PATHOL LAB MED, V113, P684
[7]   CLINICAL FLOW CYTOMETRIC RETICULOCYTE ANALYSIS [J].
DAVIS, BH ;
BIGELOW, NC .
PATHOBIOLOGY, 1990, 58 (02) :99-106
[8]   CIRCULATING RETICULOCYTES IN NORMAL ADULTS AS DETERMINED BY NEW METHYLENE BLUE METHOD [J].
DEISS, A ;
KURTH, D .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1970, 53 (04) :481-&
[9]   EVALUATION OF RETICULOCYTE COUNTS BY FLOW-CYTOMETRY IN A ROUTINE LABORATORY [J].
FERGUSON, DJ ;
LEE, SF ;
GORDON, PA .
AMERICAN JOURNAL OF HEMATOLOGY, 1990, 33 (01) :13-17
[10]   AUTOMATED RETICULOCYTE COUNTING BY IMAGE-ANALYSIS AND FLOW-CYTOMETRY [J].
HACKNEY, JR ;
CEMBROWSKI, GS ;
PRYSTOWSKY, MB ;
KANT, JA .
LABORATORY MEDICINE, 1989, 20 (08) :551-555