ACCURACY OF LATERAL DUAL-ENERGY X-RAY ABSORPTIOMETRY FOR THE DETERMINATION OF BONE-MINERAL CONTENT IN THE THORACIC AND LUMBAR SPINE - AN INVITRO STUDY

被引:25
作者
EDMONDSTON, SJ
SINGER, KP
PRICE, RI
BREIDAHL, PD
机构
[1] CURTIN UNIV, SCH PHYSIOTHERAPY,DEPT PHYSIOTHERAPY,SELBY ST, SHENTON PK 6008, PERTH, WA, AUSTRALIA
[2] ROYAL PERTH HOSP, DEPT RADIOL, PERTH, WA 6001, AUSTRALIA
[3] SIR CHARLES GAIRDNER HOSP, DEPT ENDOCRINOL & DIABET, PERTH, WA, AUSTRALIA
关键词
D O I
10.1259/0007-1285-66-784-309
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Lateral scanning with dual energy X-ray absorptiometry (DXA) was undertaken on 153 thoracic and lumbar vertebral bodies from nine individuals (mean age 67; range 34-92 years) to establish their bone mineral content (BMC). All specimens were subsequently de-fatted and ashed to compare ash weight with BMC of the T1 to L5 segments. Linear regression analysis indicated a higher correlation between BMC and measured ash weights for each vertebral column (range: r = 0.97-0.99), with error analysis for all vertebrae tested showing a standard error of 0.40 g, or 10.8%. For the L2-4 segments the accuracy error was 0.50 g or 7.9%. In both cases there was a mean trend towards underestimation of ash weight. Mean BMC values of the T1 to T5 segments were similar before a progressive increase caudally. These data affirm the utility of DXA for determining bone mineral content in the vertebral column across wide ranges of age, and segmental and bone density.
引用
收藏
页码:309 / 313
页数:5
相关论文
共 21 条
  • [21] WHITE AA, 1977, CLIN ORTHOP RELAT R, P8