ESTIMATION OF INSITU ULTRASOUND EXPOSURE DURING OBSTETRIC EXAMINATIONS

被引:12
作者
RAMNARINE, KV
NASSIRI, DK
PEARCE, JM
JOSEPH, AEA
PATEL, RH
VARMA, TR
机构
[1] ST GEORGE HOSP,DEPT OBSTET & GYNAECOL,LONDON SW17 0QT,ENGLAND
[2] ST GEORGE HOSP,DEPT DIAGNOST RADIOL,LONDON SW17 0QT,ENGLAND
关键词
ULTRASONIC SCANNING; ULTRASONICS; PREGNANCY; ULTRASONIC ATTENUATION; MATERNAL ANATOMY; FETAL MONITORING; ULTRASONIC DOPPLER; 1ST TRIMESTER; 2ND TRIMESTER; 3RD TRIMESTER;
D O I
10.1016/0301-5629(93)90104-V
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Layered tissue models are developed to estimate in situ ultrasound intensity during common obstetric examinations by incorporation of measured overlying tissue thicknesses with data on the attenuation properties of tissues. Results are compared with attenuation models recommended by the United States Food and Drug Administration (FDA) and National Council on Radiation Protection and Measurements (NCRP). For abdominal and transvaginal scanning in the first trimester, fixed attenuation models based on attenuation values of 1.2 dB/MHz and 0.6 dB/MHz, respectively, are representative of worst-case exposure conditions. For second and third trimesters, a fixed attenuation value of 0.8 dB/MHz is representative of worst-case exposure conditions. A fixed attenuation value of 1.0 dB/MHz is suggested for common Doppler examinations. The study suggests that the FDA-derating factor of 0.3 dB/cm . MHz may not give a conservative estimate of in situ intensity for certain obstetric examinations.
引用
收藏
页码:319 / 329
页数:11
相关论文
共 21 条
[1]  
BANG J, 1972, ACTA PATH MICRO IM A, VA 80, P341
[2]   MEDICAL SONOGRAPHY - REPRODUCTIVE EFFECTS AND RISKS [J].
BRENT, RL ;
JENSH, RP ;
BECKMAN, DA .
TERATOLOGY, 1991, 44 (02) :123-146
[3]   FETAL DEPTH AND ULTRASOUND PATH LENGTHS THROUGH OVERLYING TISSUES [J].
CARSON, PL ;
RUBIN, JM ;
CHIANG, EH .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (07) :629-639
[4]   NCRP-AIUM UNIFORM ABSORPTION MODELS [J].
CARSTENSEN, EL .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 :31-34
[5]   INVIVO FETAL ULTRASOUND EXPOSIMETRY [J].
DAFT, CMW ;
SIDDIQI, TA ;
FITTING, DW ;
MEYER, RA ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (06) :501-505
[6]   OUTPUT DATA FROM EUROPEAN STUDIES [J].
DUCK, FA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 :61-64
[7]   TRENDS IN DIAGNOSTIC ULTRASOUND EXPOSURE [J].
DUCK, FA ;
MARTIN, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (11) :1423-1432
[8]  
DUCK FA, 1986, BR MED ULTRASOUND SO, V10, P12
[9]   COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES .2. [J].
GOSS, SA ;
JOHNSTON, RL ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1980, 68 (01) :93-108
[10]   COMPREHENSIVE COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES [J].
GOSS, SA ;
JOHNSTON, RL ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1978, 64 (02) :423-457